Perioperative Noninvasive Blood Pressure Monitoring
1From the Department of Anesthesiology, University of Utah School of Medicine, Salt Lake City, Utah.
Abstract:
The most commonly monitored variable for perioperative hemodynamic management is blood pressure. Several indirect noninvasive blood pressure monitoring techniques have been developed over the last century, including intermittent techniques such as auscultation (Riva-Rocci and Korotkoff) and oscillometry (Marey) and continuous techniques. With the introduction of automated noninvasive blood pressure devices in the 1970s, the oscillometric technique quickly became and remains the standard for automated, intermittent blood pressure measurement. It tends to estimate more extreme high and low blood pressures closer to normal than what invasive measurements indicate. The accuracy of the oscillometric maximum amplitude algorithm for estimating mean arterial pressure is affected by multiple factors, including the cuff size and shape, the shape of the arterial compliance curve and arterial pressure pulse, and pulse pressure itself. Additionally, the technique typically assumes a consistent arterial compliance and arterial pressure pulse, thus changes in arterial compliance and arrhythmias that lead to variation in the pressure pulse can affect accuracy. Volume clamping, based on the Penaz principle, and arterial tonometry provide continuous tracking of the arterial pressure pulse. The ubiquitous use of blood pressure monitoring is in contrast with the lack of evidence for optimal perioperative blood pressure targets.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Cardiomyopathy VII: Pre and Post Operative Nursing Management
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...