Usability study of pH strips for nasogastric tube placement
Simone Borsci1, Peter Buckle1, Jeremy Huddy1
1National Institute for Health Research Diagnostic Evidence Cooperative of London, Faculty of Medicine, Department of Surgery & Cancer, Imperial College, London, United Kingdom.
Aims:
(1) To model the process of use and usability of pH strips (2) to identify, through simulation studies, the likelihood of misreading pH strips, and to assess professional's acceptance, trust and perceived usability of pH strips.
Methods:
This study was undertaken in four phases and used a mixed method approach (an audit, a semi-structured interview, a survey and simulation study). The three months audit was of 24 patients, the semi-structured interview was performed with 19 health professionals and informed the process of use of pH strips. A survey of 134 professionals and novices explored the likelihood of misinterpreting pH strips. Standardised questionnaires were used to assess professionals perceived usability, trust and acceptance of pH strip use in a simulated study.
Results:
The audit found that in 45.7% of the cases aspiration could not be achieved, and that 54% of the NG-tube insertions required x-ray confirmation. None of those interviewed had received formal training on pH strips use. In the simulated study, participants made up to 11.15% errors in reading the strips with important implications for decision making regarding NG tube placement. No difference was identified between professionals and novices in their likelihood of misinterpreting the pH value of the strips. Whilst the overall experience of usage is poor (47.3%), health professionals gave a positive level of trust in both the interview (62.6%) and the survey (68.7%) and acceptance (interview group 65.1%, survey group 74.7%). They also reported anxiety in the use of strips (interview group 29.7%, survey group 49.7%).
Conclusions:
Significant errors occur when using pH strips in a simulated study. Manufacturers should consider developing new pH strips, specifically designed for bedside use, that are more usable and less likely to be misread.
Related Concept Videos
Enteral Nutrition I: Orogastric and Nasogastric Feeding
Orogastric (OG) and nasogastric (NG) feeding are two standard methods used for enteral nutrition. Enteral nutrition is often preferred over...
Enteral Nutrition II: Nasointestinal and Gastrostomy Feeding
Nasointestinal Feeding
Nasointestinal feeding involves placing a tube...
Suctioning the Nasopharyngeal Airway
Equipment Required
Endoscopic Procedures I: Esophagogastroduodenoscopy
During an EGD, the endoscope can be used to:
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Upper GI Series: Barium Swallow
Purpose and Procedure
Patients undergoing this procedure ingest a liquid containing barium sulfate with a chalky...