Background:
The aim of the current study was to evaluate the commutability of commercial control materials and human serum pools and to investigate the suitability of the materials for the external quality assessment (EQA) of serum apolipoprotein A-I (apo A-I) measurement.
Methods:
The Clinical and Laboratory Standards Institute (CLSI) EP14-A3 protocol was used for the commutability study. Apo A-I concentrations in two levels of commercial control materials used in EQA program, two fresh-frozen human serum pools (FSPs) and two frozen human serum pools prepared from residual clinical specimens (RSPs) were measured along with 50 individual samples using nine commercial assays. Measurement results of the 50 individual samples obtained with different assays were pairwise analyzed by Deming regression, and 95% prediction intervals (PIs) were calculated. The commutability of the processed materials was evaluated by comparing the measurement results of the materials with the limits of the PIs.
Results:
The FSP-1 was commutable for all the 36 assay pairs, and FSP-2 was commutable for 30 pairs; RSP-1 and RSP-2 showed commutability for 27/36 and 22/36 assay pairs, respectively, whereas the two EQA materials were commutable only for 4/36 and 5/36 assay pairs, respectively.
Conclusions:
Non-commutability of the tested EQA materials has been observed among current apo A-I assays. EQA programs need either to take into account the commutability-related biases in the interpretation of the EQA results or to use more commutable materials. Frozen human serum pools were commutable for most of the assays.