Measurement of contact angles at room temperature in high magnetic field
Chuanjun Li1, Yang Cao1, Rui Guo1
1State Key Laboratory of Advanced Special Steel and Shanghai Key Laboratory of Advanced Ferrometallurgy and School of Materials Science and Engineering, Shanghai University, Shanghai 200072, China.
Abstract:
The contact angle (CA) goniometer adaptable to a superconducting magnet was developed based on the sessile drop method. The goniometer mainly consisted of the sampling system, the supporting system, and the image acquisition system. Some improvements were taken to avoid the effects of the magnetic field (MF) on the CA measurement. As an example, the CAs of water on two substrates of silica and polymethyl methacrylate (PMMA) were measured using the goniometer. The results with and without a MF showed a good repeatability and reliability. Additionally, the MF was found to reduce the CA of water, which probably stemmed from the change of the surface tension in the MF. The CA goniometer will become an important tool which is used to study the wettability of liquids on a solid in the MF.
Related Concept Videos
Magnetic Fields
A magnetic field is defined by the force that a charged particle experiences...
Magnetostatic Boundary Conditions
Magnetic Flux
Suppose a surface is divided into elements of area dA. For each element, the component of the magnetic field that is normal to the...
Contact Angle
The adhesive force is the molecular force between molecules of different materials, that is, between the molecules of the solid and the liquid. The cohesive...
Magnetic Field Of A Current Loop
Ferromagnetism