Related Experiment Video
Updated: Feb 17, 2026

Evaluating Primary Blast Effects In Vitro
Published on: September 18, 2017
Evaluation of an experimental electrohydraulic discharge device for extracorporeal shock wave lithotripsy: Pressure
Guangyan Li1, Bret A Connors2, Ray B Schaefer3
1School of Physics, Northeast Normal University, Changchun, 130024, People's Republic of China.
Abstract:
In this paper, an extracorporeal shock wave source composed of small ellipsoidal sparker units is described. The sparker units were arranged in an array designed to produce a coherent shock wave of sufficient strength to fracture kidney stones. The objective of this paper was to measure the acoustical output of this array of 18 individual sparker units and compare this array to commercial lithotripters. Representative waveforms acquired with a fiber-optic probe hydrophone at the geometric focus of the sparker array indicated that the sparker array produces a shock wave (P+ ∼40-47 MPa, P- ∼2.5-5.0 MPa) similar to shock waves produced by a Dornier HM-3 or Dornier Compact S. The sparker array's pressure field map also appeared similar to the measurements from a HM-3 and Compact S. Compared to the HM-3, the electrohydraulic technology of the sparker array produced a more consistent SW pulse (shot-to-shot positive pressure value standard deviation of ±4.7 MPa vs ±3.3 MPa).