Related Experiment Video
Updated: Feb 17, 2026

Potentiodynamic Corrosion Testing
Published on: September 4, 2016
Bead debonding from a modern cementless total hip femoral component with concomitant taper corrosion
Adam I Edelstein1, Duy L Phan2, Scott M Sporer2,3
1Department of Orthopaedic Surgery, Northwestern University, Feinberg School of Medicine, Chicago, IL, USA.
Abstract:
We report a case of bead shedding from a cylindrical extensively porous-coated cementless femoral component with concomitant taper corrosion at the modular head-neck junction of a metal-on-polyethylene total hip prosthesis. The patient presented with chronic thigh pain 4 years after primary total hip arthroplasty, and radiographs revealed significant osteolysis and metallic debris around the femoral stem. Intraoperatively, the patient had a grossly loose femoral component with debonding of sintered beads from the femoral stem, as well as evidence of taper corrosion. We identify a failure of a modern beaded femoral component in conjunction with taper corrosion.
Related Concept Videos
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Hydration of Cement
Mortar Joint Deterioration in Masonry
The...