Purpose:
In this work, the mechanical behavior of two block copolymers for biomedical applications is studied with particular regard to the effects induced by a steam sterilization treatment that biomedical devices usually undergo in healthcare facilities. This investigation is aimed at describing the elasto-plastic behavior of the stress-strain response, determining a functional dependence between material constitutive parameters, to obtain an optimal constitutive model.
Methods:
The mechanical properties of these polymers are analyzed through uniaxial tensile tests, before and after the sterilization process. The effect of sterilization on the mechanical behavior is evaluated. The Ramberg-Osgood model is used to describe the elasto-plastic behavior of the stress-strain response.
Results:
Data from uniaxial tensile tests are discussed in the light of previous data on the same polymeric materials, in order to highlight the correlation between physicochemical and mechanical properties variation. The material constitutive parameters are determined and the functional dependence between them is found, thus enabling an optimal constitutive model to be obtained.
Conclusions:
The effect of sterilization on the material constitutive parameters is studied, to evaluate the suitability of the model in describing the mechanical response of biomedical polymer before and after sterilization treatment. The same approach can be applied to other biomaterials, under various tensile tests, and for several processes that induce variation in mechanical properties.
More Related Videos
09:37Fabrication of a Bioactive, PCL-based "Self-fitting" Shape Memory Polymer Scaffold
Published on: October 23, 2015
07:15A Testing Platform for Durability Studies of Polymers and Fiber-reinforced Polymer Composites under Concurrent Hygrothermo-mechanical Stimuli
Published on: December 11, 2014
Related Concept Videos
Polymer Classification: Stereospecificity
Polymer Classification: Architecture
Plastic Behavior
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...