Related Experiment Video
Updated: Feb 17, 2026

Manufacture of a Multi-Purpose Low-Cost Animal Bench-Model for Teaching Tracheostomy
Published on: May 18, 2019
A randomized cross-over study comparing surgical cricothyrotomy techniques by combat medics using a synthetic cadaver
Steven G Schauer1, Jessie Renee D Fernandez2, Jamie L Roper3
1US Army Institute of Surgical Research, JBSA Fort Sam Houston, TX, United States; 59th Medical Wing, JBSA Lackland, TX, United States; Brooke Army Medical Center, JBSA Fort Sam Houston, TX, United States.
This study found no significant differences in cricothyrotomy success rates or time to cannulation between the standard surgical technique, Control-Cric™ System, and QuickTrach II™ device. The QuickTrach II™ was preferred by most participants.
Area of Science:
- Emergency Medicine
- Surgical Procedures
- Medical Device Technology
Background:
- Cricothyrotomy is a critical airway procedure with a high complication rate.
- Novel devices like Control-Cric™ and QuickTrach II™ aim to improve cricothyrotomy outcomes.
- Comparison of new devices against standard surgical techniques is essential.
Purpose of the Study:
- To compare the efficacy of the Control-Cric™ System and QuickTrach II™ against standard open surgical cricothyrotomy.
- To evaluate time to successful cannulation and first-attempt success rates.
- To assess user preference for the tested cricothyrotomy devices.
Main Methods:
- A randomized crossover study was conducted with US Army combat medics.
- Participants used a synthetic cadaver model to perform cricothyrotomy with three techniques: standard, Control-Cric™, and QuickTrach II™.
- Primary outcome: time to successful cannulation; Secondary outcome: first-attempt success; User preference survey.
Main Results:
- Mean cannulation times were comparable across all three methods (Standard: 51.0s, QuickTrach II™: 39.8s, Control-Cric™: 53.6s).
- Cannulation failure rates (Standard: 6.2%, QuickTrach II™: 13.9%, Control-Cric™: 18.5%) and first-pass success rates (93.4%, 91.1%, 88.7%) showed no significant differences.
- A majority of participants (52.3%) preferred the QuickTrach II™ device.
Conclusions:
- No statistically significant differences were found in time to successful cannulation or first-pass success among the three cricothyrotomy techniques.
- While not statistically significant, the QuickTrach II™ device demonstrated a trend towards faster cannulation and was preferred by users.
- Further research may explore factors influencing user preference and long-term device efficacy.
More Related Videos
04:43Endotracheal Intubation via Tracheotomy and Subsequent Thoracotomy in Rats for Non-Survival Applications
Published on: March 15, 2024
09:10Heterotopic and Orthotopic Tracheal Transplantation in Mice used as Models to Study the Development of Obliterative Airway Disease
Published on: January 20, 2010