Related Experiment Video
Updated: Feb 17, 2026

06:04
Systematic Hearing Performance Evaluation Process for Adolescents with Cochlear Implantation at Early Ages
Published on: March 24, 2023
874
Outcome evaluation on cochlear implant users with residual hearing
Nicole Neben1, Andreas Buechner2, Mark Schuessler2
1a Cochlear Deutschland GmbH & Co. KG , Karl-Wiechert-Allee 76A, Hannover 30625 , Germany.
Cochlear Implants International
|December 8, 2017
Summary
The CP900 series sound processor with Hybrid Hearing demonstrated superior performance compared to the Freedom Hybrid, especially in background noise. This new processor is recommended for cochlear implant users with residual hearing.
Area of Science:
- Audiology
- Biomedical Engineering
- Otorhinolaryngology
Background:
- Hybrid Hearing combines acoustic and electric stimulation for cochlear implant (CI) users.
- Assessing advanced sound processors is crucial for optimizing hearing outcomes.
- Evaluating upgrades for experienced CI users with residual hearing is important.
Purpose of the Study:
- To compare the performance of the Freedom™ Hybrid™ and CP900 series sound processors with Hybrid Hearing.
- To assess the feasibility of upgrading experienced electric-only CI users to Hybrid Hearing.
Main Methods:
- A single-centre prospective, non-inferiority study with a randomized AABB cross-over design.
- Two groups were used: Group 1 compared two Hybrid Hearing systems; Group 2 compared electric-only vs. Hybrid Hearing in experienced CI users.
- Speech perception tests and patient-reported outcomes were utilized.
Main Results:
- The CP900 series processor showed equivalent or superior performance, with a statistically significant improvement in background noise (1.87 dB).
- Speech understanding in quiet showed a trend towards better performance with the CP900 series (5% improvement).
- The feasibility group showed an average benefit of 0.54 dB in background noise with Hybrid Hearing vs. electric-only.
Conclusions:
- The CP900 series sound processor with Hybrid Hearing is effective for participants with residual hearing.
- Positive outcomes include improved speech understanding and subjective hearing.
- The CP900 series is recommended for current electric-only CI users on a case-by-case basis.