Electron beam collimation with a photon MLC for standard electron treatments
S Mueller1, M K Fix1, D Henzen1
1Division of Medical Radiation Physics and Department of Radiation Oncology, Inselspital, Bern University Hospital, and University of Bern, Bern, Switzerland.
Photon multileaf collimators (pMLC) can replace traditional cut-outs for electron beam collimation, improving efficiency and enabling advanced radiotherapy techniques. This study validates pMLC electron beams, showing comparable treatment plan quality to cut-out based plans.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Technology
Background:
- Standard electron treatments use patient-specific cut-outs for beam collimation.
- Advanced techniques like modulated electron radiotherapy (MERT) and mixed beam radiotherapy (MBBR) require more flexible collimation methods.
Purpose of the Study:
- To evaluate the dosimetric feasibility of using a photon multileaf collimator (pMLC) for electron beam collimation in standard radiotherapy treatments.
- To compare treatment plan quality and organ at risk (OAR) doses between pMLC-collimated and traditional cut-out-collimated electron beams.
Main Methods:
- A Monte Carlo model for pMLC-shaped electron beams was extended for extended source-to-surface distances (SSDs) and validated against measurements.
- pMLC-based and cut-out-based treatment plans were created for various clinical cases (breast boost, sternum, testis, parotid).
- Treatment plans were evaluated for target dose homogeneity and OAR mean doses.
Main Results:
- Monte Carlo simulations and measurements showed agreement within 3% of maximal dose or 2 mm distance to agreement.
- pMLC-based plans achieved comparable target dose homogeneity and OAR doses to cut-out-based plans.
- Slight increases in OAR mean dose were observed for pMLC plans in specific configurations (lateral OAR proximity, extended SSD with 6 MeV).
Conclusions:
- Photon multileaf collimators (pMLC) can effectively replace cut-outs for electron beam collimation in standard radiotherapy.
- Accurate dose distributions can be achieved with pMLC collimation, offering similar or improved treatment plan quality.
- This facilitates more efficient standard electron treatments and advanced techniques like MERT and MBRT.