Development of an LCD-Based Visual Field System.
Jin Ho Joo1, Jihyoung Lee2, Heecheon You2
1Department of Ophthalmology, Kyung Hee University Hospital at Gangdong, Kyung Hee University School of Medicine, Seoul, Korea.
Journal of Korean Medical Science
|December 8, 2017
Summary
The LCD-based visual field testing system (LVF) demonstrates diagnostic effectiveness comparable to the Humphrey Field Analyzer II-750i (HFA) for early glaucoma detection. Both systems show strong correlations in key metrics, supporting LVF
Area of Science:
- Ophthalmology
- Medical Devices
- Diagnostic Imaging
Background:
- Assessing the diagnostic accuracy of novel visual field testing technologies is crucial for improving patient outcomes.
- The Humphrey Field Analyzer II-750i (HFA) is a widely used standard for visual field testing.
- Evaluating alternative systems like the LCD-based visual field testing system (LVF) is essential for advancing diagnostic capabilities.
Purpose of the Study:
- To compare the diagnostic effectiveness of an LCD-based visual field testing system (LVF) against the standard HFA.
- To evaluate the performance of LVF and HFA in differentiating normal and glaucomatous eyes using key perimetric indices.
Main Methods:
- A randomized controlled crossover study involving 202 normal and 128 glaucomatous eyes.
- Utilized both LVF and HFA systems for visual field testing in all participants.
- Compared results based on Mean Deviation (MD), Pattern Standard Deviation (PSD), and Area Under the ROC Curve (AUC).
Main Results:
- Significant correlations were observed between LVF and HFA for both MD (r=0.342 normal, r=0.796 glaucoma) and PSD (r=0.363 normal, r=0.828 glaucoma).
- High AUC values indicated strong discriminatory power for both systems: MD (LVF 0.786, HFA 0.868) and PSD (LVF 0.913, HFA 0.932).
Conclusions:
- The LVF system demonstrates comparable competence to the HFA for visual field testing.
- LVF shows promise as an effective tool for the early detection of glaucoma.
Keywords:
Area Under the Receiver Operating Characteristics CurveLCD-based Visual Field Testing SystemMean DeviationPattern Standard DeviationStandard Automated Perimetry