Acute clouding of trifocal lens during implantation: a case report
Qian Liu1, Suhua Zhang2, Xiaogang Wang1
1Shanxi Eye Hospital, No. 100 Fudong Street, Taiyuan, Shanxi, 030002, People's Republic of China.
Background:
Intraoperative IOLs clouding of several kinds of hydrophilic acrylic intraocular lenses (IOLs) have been reported due to temperature changes. This phenomenon reported previously occurred in cold countries and during the winter months. However, no clinical case was reported about trifocal IOL opacification during operation. We report a case in which acute opacification of the optical region occurred simultaneously when AT LISA tri 839mp(Carl Zeiss) was implanted into the eye.
Case Presentation:
A 79-year-old woman with a cortex and nucleus cataract was scheduled to undergo right eye phacoemulsification assisted by femtosecond technique. The trifocal lens (AT LISA tri 839mp), which is made of hydrophilic acrylic (25%) with hydrophobic surface properties, was chosen for implantation. As the IOL was implanted into the eye, it became cloudy immediately. Then it was replaced by another AT LISA tri 839mp, which was transferred from lens company outside, the same phenomenon was observed. These two lenses underwent the same temperature fluctuation from cold outside to operating room. Finally, a ZCB00 (Allergan) was implanted.
Conclusions:
The acute intraoperative clouding of trifocal lens(AT LISA tri 839mp) results from fluctuation of temperature should be noticed.