Accuracy, Precision, and Trending of 4 Pulse Wave Analysis Techniques in the Postoperative Period
Martin Geisen1, Michael T Ganter2, Sonja Hartnack3
1Institute of Anaesthesiology and Intensive Care Medicine, Triemli Hospital Zurich, Zurich, Switzerland.
Objective:
The aim of this study was to analyze the accuracy, precision, and trending ability of the following 4 pulse wave analysis devices to measure continuous cardiac output: PiCCO2 ([PCCO]; Pulsion Medical System, Munich, Germany); LiDCORapid ([LCCO]; LiDCO Ltd, London, UK); FloTrac/Vigileo ([FCCO]; Edwards Lifesciences, Irvine, CA); and Nexfin ([NCCO]; BMEYE, Amsterdam, The Netherlands).
Design:
Prospective, observational clinical study.
Setting:
Intensive care unit of a single-center, teaching hospital.
Participants:
The study comprised 22 adult patients after elective coronary artery bypass surgery.
Interventions:
Three measurement cycles were performed in all patient durings their immediate postoperative intensive care stay before and after fluid loading. Hemodynamic measurements were performed 5 minutes before and immediately after the administration of 500 mL colloidal fluid over 20 minutes.
Measurements And Main Results:
PCCO, LCCO, FCCO, and NCCO were assessed and compared with cardiac output derived from intermittent transpulmonary thermodilution (ICO). One hundred thirty-two matched sets of data were available for analysis. Bland-Altman analysis using linear mixed effects models with random effects for patient and trial revealed a mean bias ±2 standard deviation (%error) of -0.86 ± 1.41 L/min (34.9%) for PCCO-ICO, -0.26 ± 2.81 L/min (46.3%) for LCCO-ICO, -0.28 ± 2.39 L/min (43.7%) for FCCO-ICO, and -0.93 ± 2.25 L/min (34.6%) for NCCO-ICO. Bland-Altman plots without adjustment for repeated measurements and replicates yielded considerably larger limits of agreement. Trend analysis for all techniques did not meet criteria for acceptable performance.
Conclusions:
All 4 tested devices using pulse wave analysis for measuring cardiac output failed to meet current criteria for meaningful and adequate accuracy, precision, and trending ability in cardiac output monitoring.
Related Concept Videos
Assessment of apical radial pulse
The A-R pulse assessment involves simultaneous evaluation of the apical and radial pulses. When the apical and radial pulse rates vary, this assessment helps identify a pulse deficit.
Pre-Procedural Preparation
Special considerations while measuring pulse
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse
The pulse serves as a clinical...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: