A Black Swan in clinical laboratory practice: the analytical error due to interferences in immunoassay methods
Aldo Clerico1, Lucia Belloni2, Cinzia Carrozza3
1Laboratory of Cardiovascular Endocrinology and Cell Biology, Department of Laboratory Medicine, Fondazione CNR Toscana G. Monasterio, Scuola Superiore Sant'Anna, Via Trieste 41, 56126 Pisa, Italy.
Abstract:
It is well known that the results of immunoassay methods can be affected by specific or non-specific interferences, ranging from 0.4% to 4.0%. The presence of interference may greatly compromise the accuracy of immunoassay analyses causing an error in the measurement, producing false-positive or false-negative results. From a clinical point of view, these analytical errors may have serious implications for patient care because they can cause misdiagnosis or inappropriate treatment. Unfortunately, it is a very difficult task to identify the irregular analytical errors related to immunoassay methods because they are not detectable by normal laboratory quality control procedures, are reproducible within the test system, may be clinically plausible and are relatively rare. The first line of defense against erroneous results is to use in laboratory practice only immunoassay systems with the highest level of robustness against interference. The second line of defense is always taking into account the possibility of interference in immunoassay results. A correct approach should be addressed on identification of samples at high risk of interference. The attainment of this goal requires a critical review of the test result in relation to patient's clinical conditions and literature data, taking into account the analytical characteristics of the immunoassay system. The experts in immunoassay systems should make every effort to find some specific and reliable quality indicators for irregular analytical errors in order to better detect and monitor erroneous immunoassay results due to specific or non-specific interferences.
Related Concept Videos
Systematic Error: Methodological and Sampling Errors
Sampling errors originate from improper sampling methods or the wrong sample population. These errors can be minimized by refining the sampling strategy. Defective instruments or faulty calibrations are the sources of instrumental...
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Errors occurring during blood pressure monitoring
Several factors...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Data Validation
Key parameters for method validation include: