Endoscopic Ultrasound-Guided Biliary Drainage: Endoscopic Ultrasound-Guided Hepaticogastrostomy in Malignant Biliary Obstruction
Published on: March 25, 2022
Endoscopic ultrasound-guided biliary drainage using a newly designed metal stent with a thin delivery system: a
Kosuke Minaga1, Masayuki Kitano2, Masahiro Itonaga3
1Department of Gastroenterology and Hepatology, Kindai University Faculty of Medicine, Osaka-Sayama, Japan.
Purpose:
This study was designed to evaluate the feasibility and safety of a newly designed self-expandable metal stent for endoscopic ultrasound-guided biliary drainage (EUS-BD) when it was delivered via three different stent delivery systems: a 7.5Fr delivery catheter with a bullet-shaped tip (7.5Fr-bullet), a 7Fr catheter with a bullet-shaped tip (7Fr-bullet), or a 7Fr catheter with a tee-shaped tip (7Fr-tee).
Methods:
This experimental study utilized a porcine model of biliary dilatation involving ten pigs. In the animal study, technical feasibility and clinical outcomes of the stent when placed with each of the delivery systems were examined. In addition, a phantom model was used to measure the resistance of these delivery systems to advancement.
Results:
Phantom experiments showed that, compared with 7Fr-bullet, 7Fr-tee had less resistance force to the advancement of the stent delivery system. EUS-BD was technically successful in all ten pigs. Fistulous tract dilation was necessary in 100% (2/2), 75% (3/4), and 0% (0/4) of the pigs that underwent EUS-BD using 7.5Fr-bullet, 7Fr-bullet, and 7Fr-tee, respectively. There were no procedure-related complications.
Conclusion:
Our newly designed metal stent may be feasible and safe for EUS-BD, particularly when delivered by 7Fr-tee, because it eliminates the need for fistulous tract dilation.