Related Experiment Video
Updated: Feb 17, 2026

Homogeneous Time-resolved Förster Resonance Energy Transfer-based Assay for Detection of Insulin Secretion
Published on: May 10, 2018
Interpreting insulin immunoassays during investigation of apparent spontaneous hypoglycaemia and insulin overdose
Julie Chemmanam1, Michelle Isaacs2, Graham R Jones3,4
1Southern Adelaide Diabetes and Endocrine Services, Flinders Medical Centre, Adelaide, South Austalia, Australia.
Abstract:
We report two cases of hypoglycaemia; one with apparently spontaneous hypoglycaemia and one with presumed insulin overdose. In both cases insulin concentration was normal when measured with the Roche immunoassay, but elevated when remeasured with the Advia Centaur immunoassay and a diagnosis of hypoglycaemia secondary to insulin analogue administration was made. These cases highlight that physicians need to understand the binding characteristics of the insulin immunoassay they use.