Intervendor Consistency and Accuracy of Left Ventricular Volume Measurements Using Three-Dimensional Echocardiography
Denisa Muraru1, Antonella Cecchetto1, Umberto Cucchini1
1Department of Cardiac, Thoracic and Vascular Sciences, University of Padua, Padua, Italy.
Background:
Intervendor consistency of left ventricular (LV) volume measurements using three-dimensional transthoracic echocardiography (3DTTE) has never been reported. Accordingly, we designed a prospective study to (1) compare head-to-head the accuracy of three three-dimensional echocardiography (3DE) systems in measuring LV volumes and ejection fraction (EF) against cardiac magnetic resonance (CMR); (2) assess the intervendor variability of LV volumes and EF; and (3) compare the accuracy of fully automated versus semiautomated (i.e., manually corrected) methods of LV endocardial delineation against CMR.
Methods:
We studied 92 patients (64% males, 52 years [95% CI, 20-83]) with a wide range of end-diastolic volumes (from 87 to 446 mL) and EFs (from 16% to 77%) using three different 3DE platforms (iE33; Vivid E9; Acuson SC2000) during the same echo study. CMR was performed within 3 ± 5 hours from the 3DE study in 35 patients.
Results:
LV volumes provided by the three 3DE systems correlated with CMR volumes: end-diastolic volume (iE33: R2 = 0.93; E9: R2 = 0.94; SC2000: R2 = 0.94), end-systolic volume (iE33: R2 = 0.93; E9: R2 = 0.95; SC2000: R2 = 0.94), and EF (iE33: R2 = 0.79; E9: R2 = 0.80; SC2000: R2 = 0.77). In the 92 patients studied, LV volumes and EFs measured with the three systems were similar. Use of fully automated endocardial border detection algorithms significantly underestimated LV volumes, and the degree of underestimation was higher with larger LV volumes.
Conclusions:
LV volumes and EFs measured with the three 3DE systems are consistent. Fully automated algorithms underestimated LV volumes. Our findings may help in the clinical interpretation of LV parameters obtained using different 3DE systems and encourage the clinical use of 3DTTE.
More Related Videos
Related Concept Videos
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies for Cardiovascular System I:Echocardiography
Indications: Echocardiography is utilized to diagnose heart failure, valve disorders, and myocardial infarction. It also assesses cardiac structures' size, shape, and motion,...