The Anomalous Influence of Spectral Resolution on Pulsed THz Time Domain Spectroscopy under Real Conditions
Vyacheslav A Trofimov1, Svetlana A Varentsova2
1Faculty of Computational Mathematics and Cybernetics, Lomonosov Moscow State University, Leninskiye Gory, Moscow 119992, Russia. vatro@cs.msu.ru.
Abstract:
We have studied the spectral resolution influence on the accuracy of the substance detection and identification at using a broadband THz pulse measured under real conditions (at a distance of more than 3 m from a THz emitter in ambient air with a relative humidity of about 50%). We show that increasing spectral resolution leads to manifestation of small-scale perturbations (random fluctuations) in the signal spectrum caused by the influence of the environment or the sample structure. Decreasing the spectral resolution allows us to exclude from consideration this small-scale modulation of the signal as well as to detect the water vapor absorption frequencies. This fact is important in practice because it allows us to increase the signal processing rate. In order to increase the detection reliability, it is advisable to decrease the spectral resolution up to values of not more than 40% of the corresponding spectral line bandwidth. The method of spectral dynamics analysis together with the integral correlation criteria is used for the substance detection and identification. Neutral substances such as chocolate and cookies are used as the samples in the physical experiment.
Related Concept Videos
NMR Spectrometers: Resolution and Error Correction
Atomic Absorption Spectroscopy: Interference
Spectral interference occurs when signals from other elements or molecules overlap with the analyte signal, falsely elevating or masking the analyte's absorbance. This interference can be corrected using Zeeman,...
Aliasing
If the sampling frequency is below the Nyquist rate, these replicas overlap, preventing the original...
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
¹H NMR: Interpreting Distorted and Overlapping Signals
As Δν decreases and the signals move closer, the doublets appear increasingly distorted. The intensities of the inner lines increase at the cost of those of the outer lines as the signals are...