Related Experiment Video
Updated: Feb 17, 2026

Computed Tomography (CT) Guided Implantation of a Totally Implantable Venous Access Port (TIVAP) through Subclavian Vein
Published on: January 13, 2026
Should We Remove the Retrievable Cook Celect Inferior Vena Cava Filter? Eight Years of Experience at a Single Center
Joohyung Son1, Miju Bae1, Sung Woon Chung1
1Department of Thoracic and Cardiovascular Surgery, Pusan National University Hospital.
Background:
The inferior vena cava filter (IVCF) is very effective for preventing pulmonary embolism in patients who cannot undergo anticoagulation therapy. However, if a filter is placed in the body permanently, it may lead to other complications.
Methods:
A retrospective study was performed of 159 patients who underwent retrievable Cook Celect IVCF implantation between January 2007 and April 2015 at a single center. Baseline characteristics, indications, and complications caused by the filter were investigated.
Results:
The most common underlying disease of patients receiving the filter was cancer (24.3%). Venous thrombolysis or thrombectomy was the most common indication for IVCF insertion in this study (47.2%). The most common complication was inferior vena cava penetration, the risk of which increased the longer the filter remained in the body (p=0.032, Exp(B)=1.004).
Conclusion:
If the patient is able to retry anticoagulation therapy and the filter is no longer needed, the filter should be removed, even if a long time has elapsed since implantation. If the filter cannot be removed, it is recommended that follow-up computed tomography be performed regularly to monitor the progress of venous thromboembolisms as well as any filter-related complications.