Advanced CT acquisition protocol with a third-generation dual-source CT scanner and iterative reconstruction
Marguerite E Faure1,2, Laurens E Swart3,4, Marcel L Dijkshoorn3
1Department of Radiology, University Hospital of Antwerp, Wilrijkstraat, 10 2650, Edegem, Belgium. marguerite.faure@uza.be.
Objectives:
Multidetector CT (MDCT) is a valuable tool for functional prosthetic heart valve (PHV) assessment. However, radiation exposure remains a concern. We assessed a novel CT-acquisition protocol for comprehensive PHV evaluation at limited dose.
Methods:
Patients with a PHV were scanned using a third-generation dual-source CT scanner (DSCT) and iterative reconstruction technique (IR). Three acquisitions were obtained: a non-enhanced scan; a contrast-enhanced, ECG-triggered, arterial CT angiography (CTA) scan with reconstructions at each 5 % of the R-R interval; and a delayed high-pitch CTA of the entire chest. Image quality was scored on a five-point scale. Radiation dose was obtained from the reported CT dose index (CTDI) and dose length product (DLP).
Results:
We analysed 43 CT examinations. Mean image quality score was 4.1±1.4, 4.7±0.5 and 4.2±0.6 for the non-contrast-enhanced, arterial and delayed acquisitions, respectively, with a total mean image quality of 4.3±0.7. Mean image quality for leaflet motion was 3.9±1.4. Mean DLP was 28.2±17.1, 457.3±168.6 and 68.5±47.2 mGy.cm for the non-contrast-enhanced (n=40), arterial (n=43) and delayed acquisition (n=43), respectively. The mean total DLP was 569±208 mGy.cm and mean total radiation dose was 8.3±3.0 mSv (n=43).
Conclusion:
Comprehensive assessment of PHVs is possible using DSCT and IR at moderate radiation dose.
Key Points:
• Prosthetic heart valve dysfunction is a potentially life-threatening condition. • Dual-source CT can adequately assess valve leaflet motion and anatomy. • We assessed a comprehensive protocol with three acquisitions for PHV evaluation. • This protocol is associated with good image quality and limited dose.
More Related Videos
09:57Four-Dimensional Computed Tomography-Guided Valve Sizing for Transcatheter Pulmonary Valve Replacement
Published on: January 20, 2022
09:32Time-Resolved, Dynamic Computed Tomography Angiography for Characterization of Aortic Endoleaks and Treatment Guidance via 2D-3D Fusion-Imaging
Published on: December 9, 2021
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Cardiac Catheterization III: Left Heart Catheterization
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT