Assessment of the scatter correction procedures in single photon emission computed tomography imaging using
Mehravar Rafati1, Hemmatollah Rouhani1, Ahmad Bitarafan-Rajabi2
1Department of Medical Physics and Radiology, Faculty of Paramedicine, Kashan University of Medical Sciences, Kashan, Iran.
Background:
Compton-scattered photons transfer incorrect spatial information. These photons are detected in used photo-peak energy window. In this study, three scatter correction procedures including dual-energy window (DEW), three energy window (TEW), and new approach were evaluated, and then the best procedure based on simulation and clinical conditions introduced.
Materials And Methods:
In this study, simulation projections and three-dimensional nonuniform rational B-spline-based Cardiac-Torso phantoms were produced by GEANT4 application for emission tomography simulation code. For clinical study, 2-day stress/rest myocardial perfusion imaging (MPI) protocol was performed with 99m Tc-sestamibi for 46 patients. Image quality parameters including contrast, signal-to-noise ratio (SNR), and relative noise of the background (RNB) were evaluated.
Results:
The simulation results showed that contrast values for DEW, TEW, and new approach were (0.45 ± 0.07, 0.5 ± 0.08, and 0.63 ± 0.09), SNR values (4.74 ± 0.94, 5.58 ± 1.08, and 6.56 ± 1.24), and RNB values (0.33 ± 0.06, 0.33 ± 0.07, and 0.33 ± 0.05), respectively. In clinical study, the contrast values for DEW, TEW, and new approach were 0.53 ± 0.03, 0.57 ± 0.07, and 0.62 ± 0.04 in rest MPI and were 0.52 ± 0.04, 0.57 ± 0.06, and 0.6 ± 0.05 in stress MPI, respectively. Moreover, for the rest images, the SNR values were 7.65 ± 1.9, 9.08 ± 2.2, and 10.2 ± 1.75 and for stress images were 7.76 ± 1.99, 9.12 ± 2.25, and 10.17 ± 2.04, respectively. Finally, RNB values for rest and stress images were 0.12 ± 0.03, 0.13 ± 0.03, and 0.13 ± 0.03, respectively.
Conclusion:
The simulation and the clinical studies showed that the new approach could be better performance than DEW, TEW methods, according to values of the contrast, and the SNR for scatter correction.
More Related Videos
12:24Computed Tomography-guided Time-domain Diffuse Fluorescence Tomography in Small Animals for Localization of Cancer Biomarkers
Published on: July 17, 2012
08:36Multi-Tracer Studies of Brain Oxygen and Glucose Metabolism Using a Time-of-Flight Positron Emission Tomography-Computed Tomography Scanner
Published on: June 7, 2024
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography