Infrared Skin Thermometry: Validating and Comparing Techniques to Detect Periwound Skin Infection
Asfandyar Mufti1, Ranjani Somayaji, Patricia Coutts
1Asfandyar Mufti, BMSc, is a Medical Student, University of Ottawa, Ontario, Canada. Ranjani Somayaji, MD, MPH, BScPT, FRCPC, is a Clinical Lecturer, Section of Infectious Disease, Department of Medicine, University of Calgary, Calgary, Alberta, Canada. Patricia Coutts, RN, is a Clinical Nurse, Mississauga Regional Wound Healing Clinic, Toronto, Ontario, Canada. R. Gary Sibbald, BSc, MD, DSc (Hons), MEd, FRCPC (Med Derm), FAAD, MAPWCA, is Professor, Medicine and Public Health, University of Toronto, Toronto, Ontario, Canada; Director, International Interprofessional Wound Care Course & Masters of Science in Community Health (Prevention & Wound Care), Dalla Lana Faculty of Public Health, University of Toronto; Past President, World Union of Wound Healing Societies; and Clinical Editor, Advances in Skin & Wound Care, Philadelphia, Pennsylvania.
Objective:
Diagnosis of wound infection can be challenging because it relies on a combination of clinical signs and symptoms that are often nonspecific. Increased periwound cutaneous temperature is a classic sign of deep and surrounding wound infection, and its quantitative measurement is one of the most reliable and valid clinical signs of deep and surrounding skin infection at the bedside. Skin surface temperature differences may be detected using commercially available noncontact infrared thermometers. However, techniques to detect temperature using noncontact infrared thermometers vary, and no studies have evaluated these methods. Two such measurement techniques include the "4-point" and "whole-wound" scanning methods. This study assessed the ability of different infrared thermometers using the aforementioned techniques to detect clinically meaningful differences in periwound cutaneous temperatures used in the diagnosis of wound infection.
Methods:
A prospective cohort study was conducted from 2015 to 2016 of consenting adult patients 18 years or older with an open wound attending a regional wound care clinic. One hundred patients with wounds underwent surface temperature measurement. Infection was not a specific inclusion criterion as the primary objective was to conduct a comparative assessment of infrared thermometry devices. Demographic data (age, height, weight, gender, and ethnicity) were also collected. Each wound was measured using 4 different noncontact infrared thermometers: Exergen DermaTemp 1001 (Exergen Corporation, Watertown, Massachusetts), Mastercraft Digital Temperature Reader (Mastercraft, Toronto, Ontario, Canada), Mastercool MSC52224-A (Mastercool Inc, Randolph, New Jersey), and Etekcity ETC-8250 Temperature Heat Pen (Etekcity, Anaheim, California). Data analysis was based on a comparison of 4 periwound skin surface temperature measurement differences (ΔT in degrees Fahrenheit) between the wound site and an equivalent contralateral control site.
Outcome Measures:
The primary outcome of the ability of each thermometer to detect a clinically significant difference in temperature was assessed with χ analysis. Paired t tests were conducted to compare the ΔT measurements of each specific thermometry device between the 2 measurement techniques. Pearson product moment correlation coefficients were calculated for the temperature ΔT for both measuring techniques for all devices to determine level of agreement. A 1-way analysis of variance was conducted to compare temperature measurements among the infrared thermometry devices.
Main Results:
There was no significant difference in the ability of each thermometer to detect a clinically meaningful difference of 3° F by either the 4-point (P = .10) or whole-wound techniques (P = .67). When a definition of 4° F was used, results were similar (4-point, P = .15; whole wound, P = .20). Comparisons among devices and techniques showed strong correlations (>0.80). Etekcity ETC-8250 and the 4-point measurement compared with the Exergen device using the whole-wound technique had a correlation coefficient of 0.72. Spearman ρ demonstrated a similarly high degree of correlation between techniques and devices, and only the Etekcity ETC-8250 device had a coefficient of 0.71 to 0.90 when compared with others. Paired t testing for each thermometry device comparing measurement techniques for raw temperatures ΔT demonstrated no significant difference (P > .05). No statistical differences among the ΔT values for the 3 different thermometers were noted when using the whole-wound method (P = .095). Similarly, no significant differences among the ΔT values were noted for the 4 different thermometers when using the 4-point method (P = .10).
Conclusions:
The results of this study demonstrate that both the 4-point and whole-wound methods of temperature acquisition using cost-efficient infrared thermometers provide accurate and similar results in clinical wound care settings.
Related Concept Videos
Assessing Body Temperature - Temporal Artery
Step 1: Perform hand hygiene and don a fresh pair of gloves to prevent cross-infection and ensure patient safety.
Step 2: Explain the procedure to the patient to establish trust. Clear communication establishes trust with the patient, ensures they understand what to expect, promotes cooperation, and enhances comfort during the procedure.
Step 3: Assess the patient's...
Assessing Body Temperature - Tympanic membrane
Step 1: Begin by practicing good hand hygiene to prevent the transmission of microorganisms.
Step 2: Turn on the thermometer and wait until the ready sign appears on the screen to ensure accurate measurement.
Step 3: Slide the probe cover in place to prevent cross-contamination.
Step 4: Instruct the patient to tilt their head to the side for comfort and check for cerumen...
Equipments Used to Measure Body Temperature
Glass-bulb Thermometer:
Glass-bulb thermometers are hollow glass tubes with a bulb tip containing liquid such as ethanol or mercury. Historically, glass bulb mercury thermometers were the standard device to measure body temperature. Today, mercury thermometers are prohibited in many countries due to the hazardous effects of mercury and the risk of exposure if the glass bulb breaks. In general,...
Temperature Measurement Sites
Oral: When assessing oral temperature, the thermometer tip should be placed under the tongue in the posterior sublingual pocket. It offers accurate readings and can be...
Assessing Body Temperature - Oral
Step 1:
Start by practicing proper hand hygiene to prevent the spread of microorganisms.
Step 2:
Take the thermometer out of the charging unit, switch it on, and wait for the ready sign.
Step 3:
Gently slide the probe cover until a click is heard. This simple action prevents cross-contamination and ensures the correct placement of the probe cover.
Step 4:
Instruct the patient to open their mouth and place...
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...