Comparison and optimization of imaging techniques in suspected physical abuse paediatric radiography
Jodie Rebecca Hampel1, Ana Pascoal2
11 Department of Radiation Protection and Imaging Physics, Barking, Havering and Redbridge University Hospitals NHS Trust, King's College London , London , UK.
Objective:
This study compares protocols in clinical use for paediatric suspected physical abuse (SPA) radiographic imaging across four National Health Service Trusts in the UK. The patient dose delivered from a SPA skeletal survey was compared between two sites using different imaging technology. Additionally, the technique in use for the abdomen anteroposterior (AP) radiographic projection was optimized at one of the participant sites.
Methods:
Retrospective data collection was performed to compare SPA protocols. Exposure details and patient dose data for SPA skeletal surveys were collected and compared. SPA skeletal surveys were performed on two anthropomorphic paediatric phantoms using two digital imaging systems. Effective dose (ED) was calculated using a dose calculator software (PCXMC v. 2.0, STUK, Helsinki, Finland) and used as a quantification of the radiation risk. For the optimization study, abdomen AP radiographs of the phantoms were acquired over a range of tube potentials (40-117 kV) for constant ED on a digital radiography (DR) system. The contrast-to-noise ratio (CNR) between "bone" and "soft tissue" in the images was measured and used as an indicator of image quality.
Results:
This study showed that there is a variation in the protocols and a range of techniques in use for SPA imaging across the four participant sites. The skeletal surveys undertaken on the newborn phantom at two sites resulted in an ED of 57 ± 3 µSv and 90 ± 4 µSv, on the DR unit and digital radiography/fluoroscopy (dRF) unit, respectively. Measurements of the abdomen AP projection achieved an improved CNR (4%) at a lower tube potential (55 kV) without increasing ED, compared with the current clinical setting (64 kV). Advances in knowledge: This study showed that an improved CNR can be achieved for newborn and 1-year-old abdomen AP radiographs using 0.1 mm copper filtration and a reduced kV (55 kV) without increasing ED.
Related Concept Videos
X-ray Imaging
Radiological Investigation I: X-ray and CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...