Related Experiment Video
Updated: Feb 16, 2026

The Use of Mixed Reality in Custom-Made Revision Hip Arthroplasty: A First Case Report
Published on: August 4, 2022
Severe Wear and Pseudotumor Formation Due to Taper Mismatch in a Total Hip Arthroplasty: A Case Report
M C Koper1, N M C Mathijssen1, F Witt2
1Department of Orthopedics, Reinier de Graaf Hospital, Reinier de Graafweg 3, 2625 AD Delft, the Netherlands. maartenkoper@outlook.com.
Case:
We describe the case of a seventy-six-year-old woman who had undergone bilateral total hip arthroplasty with Zweymüller-Metasul prostheses in 1996. After a fall sixteen years after the index procedures, radiographs suggested a taper fracture of the left total hip arthroplasty. However, revision surgery showed pseudotumor formation, with no evidence of taper fracture. Analysis of the prosthesis showed massive wear of the male stem taper caused by a mismatch between the stem taper and the head taper.
Conclusion:
This case vividly demonstrates how taper size mismatch can cause dramatic metal wear and increased release of metal ions, resulting in pseudotumor formation.
More Related Videos
09:51The Transition to an Anterior-Based Muscle Sparing Approach Improves Early Postoperative Function but is Associated with a Learning Curve
Published on: September 7, 2022
09:31Individualized Stem-positioning in Calcar-guided Short-stem Total Hip Arthroplasty
Published on: February 27, 2018