Background:
Intramedullary screw fixation is a common method of treating proximal metadiaphyseal fifth metatarsal (ie, Jones) fractures. Fatigue failure is a complication of this fixation. There are many screw designs available, including Jones fracture specific fixation, but the optimal choice of screw design is unknown. The purpose of this study was to compare the fatigue strength of Jones fracture specific screw designs as well as other commonly used screw designs. Our hypothesis was that there would be no difference in fatigue strength for Jones fracture specific screw designs at similar screw diameters.
Methods:
A study was performed to determine the fatigue bending strength of 5 different screw designs including Jones fracture specific screw designs at 3 different screw diameters. Six screws of each size and design underwent cyclic fatigue testing, and a median fatigue limit (MFL) was determined for each screw design and size.
Results:
The Stryker Asnis JFX solid 4.0-mm, 5.0-mm, and 6.0-mm screws had a higher MFL than all other screws with similar diameter tested (all P < .0001). Both Jones fracture specific screw designs (Stryker Asnis JFX solid screws and Charlotte Carolina Jones screws) had higher MFLs than the other screw designs tested.
Conclusion:
This study provides comparative fatigue strength data on larger screw diameters, which have not been previously reported. There was a statistically significant difference in screw fatigue properties at the screw diameters tested.
Clinical Relevance:
The clinical significance of this study is that it provides surgeons with fatigue strength data to aid in screw selection for Jones fracture fixation.
More Related Videos
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
06:58Method and Instrumented Fixture for Femoral Fracture Testing in a Sideways Fall-on-the-Hip Position
Published on: August 17, 2017
Related Concept Videos
Fractures: Bone Repair
Minor fractures with no bone displacement are treated by immobilizing the fractured bone using a cast or splint. However, in the case of fractures with displaced bones, the broken bones are repositioned before immobilization to ensure successful healing without deformation and loss of function. The realignment of fractured bone ends is performed through a process called reduction. If the...
Fatigue
Fatigue Strength of Concrete
Normal Strain under Axial Loading
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Frictional Forces on Screws
A jack with a square-threaded screw is a mechanical device used to lift heavy loads by applying a force at its handle. When the force is applied, the screw turns, raising the load. The screw can...