Pacing failure caused by automatic pacing threshold adjustment system
Jun Kishihara1, Shinichi Niwano1, Hidehira Fukaya1
1Department of Cardiovascular Medicine, Kitasato University School of Medicine, Sagamihara, Japan.
Abstract:
Ventricular capture management is an automatic pacing threshold adjustment algorithm that automatically measures pacing threshold through detection of the evoked response after a pacing stimulus. Although it is principally designed to save device battery under the maintenance of the patient׳s safety, we experienced a rare case with serious pacing failure due to a weakness of this algorithm. This pacing failure might be explained by a large variation in the ventricular pacing threshold depending on the atrioventricular interval and daily variation of pacing threshold and concomitant steroid use in this patient.