Accuracy of oscillometric blood pressure measurement in atrial fibrillation
Rudolf K Feenstra1, Cor P Allaart2,3, Gijs F N Berkelmans4
1VUmc School of Medical Sciences, Amsterdam.
Insights
Automated oscillometry (AO) has limited accuracy for blood pressure measurement in atrial fibrillation (AF) patients. Accuracy decreases with high blood pressure variability and heart frequency.
Area of Science:
- Cardiology
- Biomedical Engineering
Background:
- Atrial fibrillation (AF) presents challenges for accurate blood pressure (BP) monitoring.
- Automated oscillometry (AO) is a common non-invasive BP measurement technique.
Purpose of the Study:
- To evaluate the accuracy of AO for BP measurement in outpatients with AF.
- To investigate the impact of beat-to-beat BP variability and heart frequency (HF) on AO accuracy in AF patients.
Main Methods:
- Comparison of AO measurements (Welch Allyn Spot Vital Signs and Philips M3002A IntelliVue ×2) against beat-to-beat BP using the Volume Clamp Method (VCM).
- Analysis of AO accuracy stratified by tertiles of beat-to-beat BP variability and HF.
Main Results:
- AO measurements showed deviations from VCM in both AF and sinus rhythm (SR) patients.
- Accuracy of AO (both SBP and DBP) was reduced in patients with high beat-to-beat BP variability and high HF.
- Specific deviations were noted for both devices tested.
Conclusions:
- Automated oscillometry demonstrates limited accuracy for BP monitoring in AF patients.
- Inaccuracy of AO can be substantial in individual AF patients.
- High beat-to-beat BP variability and high HF significantly impair AO accuracy in AF.
Objective:
The primary aim of this study was to assess the accuracy of automated oscillometry (AO) in outpatients with atrial fibrillation (AF). The secondary aim was to explore whether AO accuracy is influenced by beat-to-beat blood pressure (BP) variability or heart frequency (HF).
Methods:
Systolic blood pressure (SBP) and diastolic blood pressure (DBP) were measured by AO and beat-to-beat BP using a validated Volume Clamp Method (VCM) technique. AO accuracy was analyzed separately in tertiles of beat-to-beat BP variability and HF.
Results:
The main study included 58 AF and 38 sinus rhythm (SR) patients in whom the Welch Allyn Spot Vital Signs (WASVS) was used. An auxiliary study in 23 AF patients used the Philips M3002A IntelliVue ×2. For AF and SR patients, respectively, SBP by WASVS deviated by +0.1 (±14.8) mmHg and -7.9 (±15.7) mmHg from VCM. WASVS-DBP was higher than VCM in AF and SR by 6.3 (±9.2) mmHg and 5.0 (±7.7) mmHg, respectively. High beat-to-beat BP variability and high HF decreased WASVS accuracy for both SBP and DBP. SBP and DBP measurements by Philips M3002A IntelliVue ×2 deviated by -6.8 (±13.2) mmHg and 9.4 (±8.1) mmHg, respectively.
Conclusion:
Overall, AO accuracy in AF is limited; in individual patients, AO inaccuracy may be considerable. AO accuracy is especially reduced in patients showing large beat-to-beat BP variability or high HF.
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(two-step method)
Errors occurring during blood pressure monitoring
Several factors...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...