Ex vivo human bile duct radiofrequency ablation with a bipolar catheter
Mustafa Atar1, Abdurrahman Kadayifci2,3, Ebubekir Daglilar1
1Division of Gastroenterology, Massachusetts General Hospital, Harvard Medical School, 3-H GI Associates, Zero Emerson Place, Blossom St., Boston, MA, 02114, USA.
Background:
Management of the primary and secondary tumors of the bile ducts still remains as a major clinical challenge. Radiofrequency (RF) ablation (RFA) of these tumors is feasible but the effect of RF energy on the human common bile duct (CBD) and surrounding tissues has not been investigated. This pilot study aimed to determine the relationship between RF energy and the depth of ablation in the normal human CBD.
Methods:
The study was performed on fresh ex vivo human biliary-pancreatic tissue which had been resected for a pancreatic cyst or mass. The study was conducted within 15 min after resection. A bipolar Habib RFA catheter was placed into the middle of the intact CBD, and three different (5, 7, 10 W) power settings were applied over a 90-s period by an RF generator. Gross and histological examinations were performed. The depth of coagulation necrosis in CBD and the effect of RFA on CBD wall and surrounding pancreas tissue were determined by microscopic examination.
Results:
The study included eight tissue samples. 5 W power was applied to three sites and RFA caused only focal epithelial necrosis limited to the CBD mucosa. 7 and 10 W were applied to five sites and coagulation necrosis occurred in all cases. Microscopically, necrosis was transmural, involved accessory bile duct glands, and extended to the surrounding pancreatic tissue in four of these cases. Macroscopically, RFA resulted in circumferential white-yellowish color change extending approximately 2 cm of the CBD.
Conclusion:
Bipolar RF energy application with 5 W resulted in limited ablation on CBD wall. However, 7 and 10 W generated tissue necrosis which extended through the CBD wall and into surrounding pancreas tissue. Endoscopic biliary RFA is an effective technique for local biliary tissue ablation but the use of high energy may injure surrounding tissue.
More Related Videos
04:20A Saline/Bipolar Radiofrequency Energy Device As an Adjunct for Hemostasis in Solid Organ Injury/Trauma
Published on: July 28, 2020
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017