On accuracy and precision of flowmeters used for oxygen therapy in a veterinary teaching hospital
Morgan L Murphy1, David S Hodgson1, Nora M Bello2
1Department of Clinical Sciences, College of Veterinary Medicine, Kansas State University, Manhattan, KS, USA.
Objective:
To determine the accuracy and precision of flowmeters used for oxygen therapy in a veterinary teaching hospital.
Study Design:
An observational study.
Methods:
A total of 50 flowmeters used for oxygen therapy were evaluated using Defender 530 gas flow analyzers to measure flow. For each flowmeter, a minimum of seven flow settings were tested in random order and in triplicate. Flow measured at ambient conditions was converted to standardized flow specifications (21.1 °C and 760 mmHg) and analyzed using general linear mixed models. Flowmeters were considered accurate at a given flow setting when the targeted mean flow was within the corresponding 95% confidence interval. Precision of flow was characterized based on the magnitude of variance component estimates.
Results:
Flowmeters of 1.0, 3.5 and 8.0 L minute-1 were considered accurate across flow settings corresponding to their capacity range. Flowmeters of 7.0 and 15.0 L minute-1 were accurate at flow settings ≤2.0 L minute-1. For flow settings ≥3.0 L minute-1, average oxygen flow was consistently below reference values. Precision varied with the capacity of the flowmeter, ranked by decreasing precision as 1.0 > 3.5 > 8.0 > 7.0 > 15.0 L minute-1.
Conclusions And Clinical Relevance:
A flowmeter of the smallest maximum capacity within the desired flow range is more appropriate for smaller patients where accurate, precise flow delivery is needed. Although 15.0 L minute-1 flowmeters were accurate at flow settings ≤2.0 L minute-1, the graduated increments do not allow exact flow settings <0.5 L minute-1. Flowmeters of 15 L minute-1 capacity should be useful for high-flow oxygen delivery for which accuracy and precision are not critical.
More Related Videos
Related Concept Videos
Pipe Flowrate Measurement
The orifice meter is a simple,...
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Administering Oxygen by Mask
Administering oxygen by mask is a common nursing intervention that provides supplemental oxygen to patients with respiratory distress or chronic lung conditions. This procedure involves delivering oxygen at a specified rate through a face mask connected to an oxygen source.
Equipment
The equipment necessary for this procedure includes: