Quantitative Flow Imaging in Human Umbilical Vessels In Utero Using Nongated 2D Phase Contrast MRI
Uday Krishnamurthy1,2, Brijesh K Yadav1,2, Pavan K Jella1
1Department of Radiology, Wayne State University School of Medicine, Detroit, Michigan, USA.
Background:
Volumetric assessment of afferent blood flow rate provides a measure of global organ perfusion. Phase-contrast magnetic resonance imaging (PCMRI) is a reliable tool for volumetric flow quantification, but given the challenges with motion and lack of physiologic gating signal, such studies, in vivo on the human placenta, are scant.
Purpose:
To evaluate and apply a nongated (ng) PCMRI technique for quantifying blood flow rates in utero in umbilical vessels.
Study Type:
Prospective study design.
Study Population:
Twenty-four pregnant women with median gestational age (GA) 30 4/7 weeks and interquartile range (IQR) 8 1/7 weeks.
Field Strength/Sequence:
All scans were performed on a 3.0T Siemens Verio system using the ng-PCMRI technique.
Assessment:
The GA-dependent increase in umbilical vein (UV) and arterial (UA) flow was compared to previously published values. Systematic error to be expected from ng-PCMRI, in the context of pulsatile UA flow and partial voluming, was studied through Monte-Carlo simulations, as a function of resolution and number of averages.
Statistical Tests:
Correlation between the UA and UV was evaluated using a generalized linear model.
Results:
Simulations showed that ng-PCMRI measurement variance reduced by increasing the number of averages. For vessels on the order of 2 voxels in radius, partial voluming led to 10% underestimation in the flow. In fetuses, the average flow rates in UAs and UV were measured to be 203 ± 80 ml/min and 232 ± 92 ml/min and the normalized average flow rates were 140 ± 59 ml/min/kg and 155 ± 57 ml/min/kg, respectively. Excellent correlation was found between the total arterial flow vs. corresponding venous flow, with a slope of 1.08 (P = 0.036).
Data Conclusion:
Ng-PCMRI can provide accurate volumetric flow measurements in utero in the human umbilical vessels. Care needs to be taken to ensure sufficiently high-resolution data are acquired to minimize partial voluming-related errors.
Level Of Evidence:
2 Technical Efficacy Stage 1 J. Magn. Reson. Imaging 2017.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Phase Transitions
Phase Diagrams
Two-Dimensional (2D) NMR: Overview
The first step is the preparation period, during which nucleus A is excited with a radiofrequency pulse....
Inductance: Single-Phase And Three-Phase Line
Single-Phase Two-Wire Line:
A single-phase line consists of two solid cylindrical conductors, denoted as x and y. Each conductor carries phasor currents ix and iy, respectively. Given that the sum of these currents is...