Objectives:
Continuous monitoring blood culture systems are commonly used for sterile body fluid cultures. In this multicentre study, we evaluated the performance of the new-generation BacT/Alert Virtuo system compared to the BacT/Alert 3D and conventional culture for the recovery of microorganisms from sterile body fluids.
Methods:
Peritoneal, cerebrospinal, pericardial, pleural and synovial fluids from adult patients submitted for culture were collected from three different centres. Specimens were inoculated into two bottles of the same bottle type (SA, SN, FA Plus or FN Plus) in equal volumes for simultaneous incubation in the Virtuo and 3D instruments. Each specimen was also Gram stained and seeded to solid media.
Results:
A total of 811 specimens were inoculated to 1257 bottle pairs. The Virtuo and 3D showed equivalent recovery of clinically significant microorganisms (127/155, 81.9%, vs. 126/155, 81.3%, respectively). Solid media cultures recovered fewer pathogens than either continuous monitoring system (95/155, 61.3%, p <0.001), including significantly fewer Enterobacteriaceae and enterococci. The Virtuo was significantly faster than the 3D in median time to detection of isolates from the same specimen (12.5 (range, 2.8-101.5) hours vs. 15.5 (range, 4.3-78.5) hours, p <0.001). Direct specimen Gram stain detected the eventual pathogen in 30 (26.1%) of 115 significant positive specimens.
Conclusions:
The BacT/Alert Virtuo system was equivalent to the 3D system in organism recovery from sterile body fluid culture but showed faster detection of growth as a result of design enhancements.
More Related Videos
07:31Implementation of a Real-Time Psychosis Risk Detection and Alerting System Based on Electronic Health Records using CogStack
Published on: May 15, 2020
08:18Automated Microbial Cultivation and Adaptive Evolution using Microbial Microdroplet Culture System MMC
Published on: February 18, 2022
Related Concept Videos
Composition of Body Fluids
Body Water Content and Fluid Compartments
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...
Methods of Sterilization I: Physical Methods
Steam sterilization uses non-toxic, low-cost moist heat in the form of saturated steam under pressure, which is fast, microbicidal, and sporicidal, and quickly warms and penetrates fabrics. Autoclaves, or steam sterilizers, expose each item to direct steam contact for a predetermined time at the necessary...
The Fluid Mosaic Model