Related Experiment Video
Updated: Feb 16, 2026

Radiation Planning Assistant - A Streamlined, Fully Automated Radiotherapy Treatment Planning System
Published on: April 11, 2018
Evaluation of a software module for adaptive treatment planning and re-irradiation
Anne Richter1, Stefan Weick2, Thomas Krieger2
1Department of Radiation Oncology, University of Wuerzburg, Josef-Schneider-Str. 11, 97080, Wuerzburg, Germany. richter_a3@ukw.de.
Background:
The aim of this work is to validate the Dynamic Planning Module in terms of usability and acceptance in the treatment planning workflow.
Methods:
The Dynamic Planning Module was used for decision making whether a plan adaptation was necessary within one course of radiation therapy. The Module was also used for patients scheduled for re-irradiation to estimate the dose in the pretreated region and calculate the accumulated dose to critical organs at risk. During one year, 370 patients were scheduled for plan adaptation or re-irradiation. All patient cases were classified according to their treated body region. For a sub-group of 20 patients treated with RT for lung cancer, the dosimetric effect of plan adaptation during the main treatment course was evaluated in detail. Changes in tumor volume, frequency of re-planning and the time interval between treatment start and plan adaptation were assessed.
Results:
The Dynamic Planning Tool was used in 20% of treated patients per year for both approaches nearly equally (42% plan adaptation and 58% re-irradiation). Most cases were assessed for the thoracic body region (51%) followed by pelvis (21%) and head and neck cases (10%). The sub-group evaluation showed that unintended plan adaptation was performed in 38% of the scheduled cases. A median time span between first day of treatment and necessity of adaptation of 17 days (range 4-35 days) was observed. PTV changed by 12 ± 12% on average (maximum change 42%). PTV decreased in 18 of 20 cases due to tumor shrinkage and increased in 2 of 20 cases. Re-planning resulted in a reduction of the mean lung dose of the ipsilateral side in 15 of 20 cases.
Conclusion:
The experience of one year showed high acceptance of the Dynamic Planning Module in our department for both physicians and medical physicists. The re-planning can potentially reduce the accumulated dose to the organs at risk and ensure a better target volume coverage. In the re-irradiation situation, the Dynamic Planning Tool was used to consider the pretreatment dose, to adapt the actual treatment schema more specifically and to review the accumulated dose.
Related Concept Videos
Sampling Plans
Random sampling is a method where each member of the population has an equal chance of being selected for the sample. It involves selecting individuals randomly, often using random number generators or lottery-type methods. For example, when analyzing the properties of a...
Planning Nursing Care I
Planning Nursing Care II
Natural Selection and Adaptation
Beyond physical adaptations,...
Adaptability of Cytoskeletal Filaments
Introduction to Innate and Adaptive Immunity
Innate immunity is the body's natural, nonspecific defense system that acts quickly to protect against pathogens. It incorporates physical barriers like skin and mucous membranes and cellular elements such as phagocytes and natural killer cells. This part of our immune system provides an immediate,...