Assessment of Needle Tip Deflection During Transrectal Guided Prostate Biopsy: Implications for Targeted Biopsies
Daniel Halstuch1,2, Jack Baniel1,2, David Lifshitz1,2
11 Division of Urology, Rabin Medical Center , Petah-Tikva, Israel .
Objectives:
To measure needle tip deflection during transrectal ultrasound (TRUS) prostate biopsy and evaluate predictors for needle tip deflection.
Materials And Methods:
Analysis of 568 prostate biopsies obtained from 51 consecutive patients who underwent a standard 12-core TRUS guided prostate biopsy. TRUS guided prostate biopsies were performed using BK flex500, with a side-fire biplane probe. Each biopsy core image was captured and clinical data were recorded prospectively. The angle between the expected trajectory of the needle and actual needle course was measured using the longitudinal view of the captured image. The distance between expected and actual needle tip was calculated. We measured median and interquartile needle tip deflection rate stratified by side and location (apex, midgland, base). Univariable and multivariable linear regressions analysis were performed.
Results:
The overall median needle tip deflection was 1.77 mm (IQR 1.35-2.47). Location did not significantly alter needle deflection measurements. On multivariable linear regression analysis, higher prostate volume (B = 0.007 95%, CI 0.004, 0.011; p < 0.001) and the right sided biopsy (B = 0.191 95%, CI 0.047, 0.336; p = 0.010) emerged as predictors of higher needle tip deflection.
Conclusions:
To the best of our knowledge this is the first study to measure needle tip deflection during TRUS guided prostate biopsies. We demonstrated that larger prostate size and biopsy side may affect the accuracy of biopsies. These results may have clinical implication to those performing targeted biopsies.
More Related Videos
06:08A Cognitive Fusion-guided Prostate Biopsy Using Multiparametric Magnetic Resonance Imaging and Transrectal Ultrasound
Published on: March 21, 2025
05:32Povidone Iodine Rectal Preparation at Time of Prostate Needle Biopsy is a Simple and Reproducible Means to Reduce Risk of Procedural Infection
Published on: September 21, 2015
Related Concept Videos
Deflection of a Beam
Singularity functions, described in an earlier lesson, are powerful mathematical tools that represent discontinuities within a function commonly encountered in structural loading...
Maximum Deflection
The maximum deflection occurs at a specific point, known as point O, where the tangent to the deflection curve is horizontal. To find point O, the slope of the tangent at any...
Targeted Cancer Therapies
There are several types of targeted therapies against...
Targets for Drug Action: Overview
Receptors are either membrane-spanning or intracellular proteins, which upon binding a ligand, get activated and transmit the signal downstream to elicit a response. Drugs bind receptors, either mimicking the action of endogenous ligands or blocking the receptor activity to bring about a modified response. Nearly 35% of approved drugs target the G...
Target Cell Response to Hormones
Notably, the cellular response can be regulated by altering the number of receptors expressed in the cell. For example, prolonged exposure to elevated hormone levels results in a gradual decline or down-regulation in the number of receptors for that specific hormone on the cell surface. Conversely, in response to low hormone levels, cells may use up-regulation, producing an...
Pharmacogenomics: Identification of New Drug Targets