Dosimetric Validation of Volumetric Modulated Arc Therapy with Three 6MV Beam-Matched Linear Accelerators
Sangaiah Ashokkumar1,2, K M Ganesh, K Ramalingam
1Research and Development center, Bharathiar University, Coimbatore, India.
Abstract:
Background: To avoid inconvenience to patients due to linear accelerator down time in busy radio-therapy departments, treatment plans can be switched between linear accelerators provided that all exhibit the same same dosimetric characteristics. In other words linear accelerators should be beam-matched. Aim: The aim of this study was to evaluate the clinical significance of beam-matching using VMAT plans. Materials and Methods: Dosimetric data with a 6MV beam from am Clinac 2100CD were taken as baseline values and other two units, a 2300CD and a Unique Performance, were factory tuned in accordance. An analysis of PDD data was performed for different field sizes to evaluate energy matching. Beam profiles for field sizes of 10×10 cm2 and 40 × 40 cm2 at depths of 1.5 cm and 10 cm were analyzed. The relative output factor and MLC dosimetric properties were compared with each machine to determine variability among the different models. Thirty patients from our database were selected, ten each for head and neck, thorax and pelvis sites. VMAT plans were created in the Eclipse treatment planning system for a Clinac 2100 CD for reference. and verification plans were created for each to compare point dose measurements. Results: The TPR 20/10 for 10 × 10 cm2 was well matched, showing no energy differences. Deviation of all point dose measurements fell within ±3%. Planar dose maps all showed greater than 95% of points with a passed area γ-value less than 1. Conclusion: Our study evaluation of beam matching with treatment planning modeling showed good agreement fior 6 MV beams across all three linear accelerators used in our clinical environment.
More Related Videos
08:39Shaping the Amplitude and Phase of Laser Beams by Using a Phase-only Spatial Light Modulator
Published on: January 28, 2019
08:27Design and Validation of a Volumetric-extrusion Bioprinter for Bioprinting of Soluble Basement Membrane Extract for Translational Research
Published on: March 28, 2025
Related Concept Videos
Arc Length Function
Sign Test for Matched Pairs
To conduct the sign test, we first calculate the differences in...
Arc Length of a Curve: Problem Solving
Reliability and Validity
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Accelerators
The effectiveness of calcium chloride can...