Development and validation of a stability-indicating HPLC method for topiramate using a mixed-mode column and charged
Eduardo Costa Pinto1,2, Mariana da Silva Gonçalves1, Lucio Mendes Cabral1
1Department of Pharmaceutics, Faculty of Pharmacy, Federal University of Rio de Janeiro, Rio de Janeiro, RJ, Brazil.
Abstract:
The analysis of topiramate in the presence of its main degradation products is challenging due to the absence of chromophore moieties and their wide range of polarity. Mixed-mode chromatography has been used in such cases because it combines two or more modes of separation. Charged aerosol detector is also an alternative since its detection is independent of optical properties and analyte ionization. This study is aimed to develop and validate two new stability-indicating methods by high-performance liquid chromatography for the main degradation products of topiramate using mixed-mode chromatography and a charged aerosol detector. Method 1 employed an Acclaim Trinity P1® column (3.0 mm × 150 mm, 2.7 μm) with a mobile phase comprising of 80% ammonium acetate buffer (20 mM, pH 4.0) and 20% methanol at a flow rate of 0.5 mL/min at 35°C. Method 2 utilized a C18 Acclaim 120® column (4.6 mm × 250 mm; 5 μm) with ACN/water (50:50) at a flow rate of 0.6 mL/min at 50°C. Validation of the two methods demonstrated excellent performance with respect to linearity, precision, accuracy, and selectivity. The limits of detection for topiramate, fructose, sulfate, sulfamate, and compound A were 2.97, 12.08, 4.02, 13.91, and 3.94 μg/mL, respectively.
Related Concept Videos
Formal Charges
RNA Stability
Reliability and Validity
Ions and Ionic Charges
In Vitro Drug Release Testing: Overview, Development and Validation
What is a Mode?
There can be more than one mode in a data set if multiple values have the same highest frequency. For instance, suppose that the Statistics exam scores of 20 students are: 50; 53; 59; 59; 63; 63; 72; 72; 72; 72; 72; 76; 78; 81; 83; 84; 84; 84; 90; 93. Here, the mode is 72, as it occurs most frequently, five times.
A data set with two modes is called bimodal. For example,...