Characterization of a neutron calibration field with 241Am-Be source using Bonner sphere spectrometer
Thiem Ngoc Le1, Hoai-Nam Tran2, Quynh Ngoc Nguyen1
1Institute for Nuclear Science and Technology, VINATOM, 179 Hoang Quoc Viet, Hanoi, Vietnam.
Abstract:
This paper presents the characterization of a neutron calibration field of a 241Am-Be source using Bonner sphere spectrometer. The characterization of the neutron field in terms of neutron ambient dose equivalent, H*(10), rates and neutron flux spectra was performed using generalized-fit and semi-empirical fit methods together with an unfolding technique using the MAXED code. Monte Carlo simulations using MCNP5 have also been conducted to qualify the characterization process. Comparison of the neutron fluxes and the H*(10) rates obtained from the various methods shows that the components of the neutron fluxes are in good agreement with a discrepancy of about 2% while the discrepancy of the direct H*(10) rates is within 3%. This good agreement among the methods indicates the reliability of the characterization process for practical calibration of neutron measuring devices. The results also suggest that the calibration should be conducted at a distance less than 200cm to ensure the contribution of the scattered component in the reading of the device less than that of the direct component by 40%.
More Related Videos
Related Concept Videos
Electric Field of a Non Uniformly Charged Sphere
Consider a non-uniformly charged sphere, for which the density of charge depends only on the distance from a point in space and not on the direction. Such a sphere has a spherically symmetrical charge distribution. Here, the electric...
IR Spectrometers
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Mass Spectrometers
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
NMR Spectrometers: Overview