A Rabbit Aortic Valve Stenosis Model Induced by Direct Balloon Injury
Published on: March 31, 2023
Avoiding S3 Valve Over-Sizing by Deployment Balloon Over-Filling: Impact on Rates of Permanent Pacemaker and Other
Tej Sheth1, Madhu K Natarajan, Catherine Kreatsoulas
1Hamilton General Hospital, 237 Barton Street East, Hamilton, Ontario, Canada, L8L 2X2. shetht@mcmaster.ca.
Objectives:
Patients with annular areas just above nominal S3 valve areas are at increased risk of over-sizing if a larger valve is implanted. We therefore evaluated the rate of permanent pacemaker (PPM) implantation associated with avoiding over-sizing by selective deployment balloon over-filling during transcatheter aortic valve replacement (TAVR) with the Sapien 3 (S3) valve.
Methods:
We included consecutive patients treated with the S3 valve from January 2016 to May 2017. We identified computed tomography annular areas where the nominally deployed valve would be over-sized by >12%-15% (areas 340-360 mm² for 23 mm valve, 420-450 mm² for 26 mm valve, 530-580 mm² for 29 mm valve) as those at highest risk for valve over-sizing. In these situations, we used the smaller valve and over-filled the deployment balloon to achieve a predicted valve area/annular area ratio of approximately 1. For annular areas >650 mm², we over-filled the 29 mm valve to achieve a similar ratio.
Results:
We evaluated 102 patients (59 males; mean age, 83.7 ± 6.5 years; mean STS score, 10.2). Over-filling of the deployment balloon was used in 35 cases (34%). We observed a post-TAVR PPM rate of 6.9% overall and 2.7% among the 75 patients without pre-TAVR right bundle-branch block (RBBB). Cases with valve over-filling vs nominal deployment had infrequent need for postdilation (14.3% vs 6.0%, respectively; P=.17) and similar postprocedure gradients (9.9 mm Hg vs 10.3 mm Hg, respectively; P=.59).
Conclusion:
A strategy to avoid S3 valve over-sizing by selective deployment balloon over-filling was associated with a low rate of PPM, especially in patients without pre-existing RBBB.
Related Concept Videos
Hemodialysis II: Procedure and Complications
Frustration and Conflict: Avoidance-Avoidance, Double-Approach Avoidance
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Impact of Groups on Groups
Pneumonia III: Complications and Assessment
Diabetes: Symptoms, Diagnosis, and Complications