CT ventilation imaging derived from breath hold CT exhibits good regional accuracy with Galligas PET
Enid M Eslick1, John Kipritidis2, Denis Gradinscak3
1Radiation Physics Laboratory, Sydney Medical School, University of Sydney, Camperdown, Australia.
Background And Purpose:
CT ventilation imaging (CTVI) derived from four dimensional CT (4DCT) has shown only moderate spatial accuracy in humans due to 4DCT image artefacts. Here we assess the accuracy of an improved CTVI using high quality exhale/inhale breath-hold CT (BHCT).
Materials And Methods:
Eighteen lung cancer patients underwent exhale/inhale BHCT, 4DCT and Galligas PET ventilation scans in a single imaging session. For each BHCT and 4DCT scan, we performed deformable image registration (DIR) between the inhale and exhale phase images to quantify ventilation using three published metrics: (i) breathing induced lung density change, CTVIDIR-HU (ii) breathing induced volume change CTVIDIR-Jac and (iii) the regional air-tissue product, CTVIHU Spatial accuracy was reported as the voxel-wise Spearman correlation r between CTVI and Galligas PET.
Results:
For BHCT-based CTVIs (N = 16), the CTVIDIR-HU, CTVIDIR-Jac and CTVIHU methods yielded mean (range) r values of 0.67 (0.52-0.87), 0.57 (0.18-0.77) and 0.49 (0.14-0.75) respectively. By comparison the 4DCT-based CTVIs (n = 14) had values of 0.32 (-0.04 to 0.51), 0.16 (-0.31 to 44) and 0.49 (0.20-0.77) respectively.
Conclusions:
High quality CT imaging is a key requirement for accurate CT ventilation imaging. The use of exhale/inhale BHCT can improve the accuracy of CTVI for human subjects.
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Radiological Investigation I: X-ray and CT
Breathing
Improving Translational Accuracy