[Experimental determination of correction factors of four detectors used in small field radiotherapy]
L Ben Salem1, A Essadok1, I Saidani1
1Unité de radiophysique, service de radiothérapie, institut Salah-Azaiz de Tunis, boulevard 9-Avril, 1006 Tunis, Tunisie.
Purpose:
The aim of this work is to determine experimentally the correction factors [Formula: see text] for four active commercial dosimeters: two microchambers and two diode detectors based on the output factor measured with radiochromic film for a radiotherapy linear accelerator equipped with circular cones.
Materials And Methods:
Initially, a radiochromic film dosimetry measurement protocol with an accuracy of 2% was developed to approach the "reference output factor". Afterwards, the corrective factors of four detectors were determined for two ionization chambers (PinPoint PTW 31016 3D, Micropoint Extradin A16) and two diodes (PTW T60017 Diode, PTW-60019 Micro-Diamond). These measurements were carried out under conical BrainLAB® collimators defining circular fields with diameters equal to 7.5mm, 10mm, 12.5mm, 15mm, 17.5mm, 20mm, 25mm, 30mm, 35mm and 45mm of a 6MV X-ray beam generated by the ClinaciX linear accelerator (Varian®). These factors are weakly dependent on the type of accelerator, whether the model and the collimation type. This allowed their comparisons with those published for the same type of detector and for an accelerator with the same index of beam quality.
Results:
The correction factors obtained experimentally were comparable in maximum deviation of 1.9% with published ones of the works using the same type of detector (mark and model) and an accelerator delivering the same beam quality for the same field size at the measurement point.
Conclusion:
The measurement protocol using the EBT3 film, which was used as a passive dosimeter to determine the "reference output factor", was validated by comparing measured and published data of active detector correction factors.
More Related Videos
Related Concept Videos
Power Factor Correction
Experimental Determination of Chemical Formula
Determining Electric Field From Electric Potential
In general, regardless of whether the electric field is uniform, it points in the direction of decreasing potential because the force on a positive...
Distance Corrections
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...