Effect of Spatial Position in the Field of View on Dimensional Changes in Cone Beam Computed Tomography
Mehrdad Panjnoush1, Yasaman Kheirandish2, Negar Zeini3
1Associate Professor, Dental Research Center, Dentistry Research Institute, Tehran University of Medical Sciences, Tehran, Iran; Department of Oral and Maxillofacial Radiology, School of Dentistry, Tehran University of Medical Sciences, Tehran, Iran.
Objectives:
This study aimed to assess the relation between dimensional changes and object location in the field of view (FOV) using cone beam computed tomography (CBCT).
Materials And Methods:
A custom-made phantom was fabricated from base plate wax. To analyze the accuracy of measurements in horizontal and longitudinal dimensions, aluminum squares (0.5 mm thickness, 10×10 mm dimensions) were constructed and placed in three levels (upper, middle, and lower) and five positions (central, right, left, anterior and posterior). This phantom was scanned using Asahi, Planmeca and NewTom CBCT systems. CBCT scans were measured three times by use of their corresponding software. Statistical analysis was performed using one-way ANOVA, post-hoc test and two-way ANOVA (P<0.05).
Results:
The differences between the mean horizontal dimensions of different systems were not significant (P=0.296). However, the differences between the mean longitudinal dimensions of different systems were significant (P=0.039). The differences between the different positions and the mean horizontal and longitudinal dimensions were significant (P<0.001, and P<0.001, respectively). The differences between the mean horizontal dimensions and different levels were not significant (P=0.51), but the differences between the mean longitudinal dimensions and different levels were significant (P<0.001). The interaction effect of level and position on the accuracy of horizontal and longitudinal measurements was significant (P<0.0001).
Conclusions:
We found statistically significant differences in most of our comparisons; however, these differences were not clinically significant. Therefore, CBCT could be an accurate device for measurement of dimensions of objects placed in different positions in the FOV.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
09:49Dual-phase Cone-beam Computed Tomography to See, Reach, and Treat Hepatocellular Carcinoma during Drug-eluting Beads Transarterial Chemo-embolization
Published on: December 2, 2013
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Field Application of Global Positioning System
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Design Example: Identifying the Locations of Monuments in the Field Using Global Positioning System Device
Econometric Views (EViews)