Fast acquisition cone-beam computed tomography: initial experience with a 10 s protocol
Visish M Srinivasan1, Gouthami Chintalapani2, Kevin Matthew Camstra1
1Department of Neurosurgery, Baylor College of Medicine, Houston, Texas, USA.
Background:
Cone-beam computed tomography (CBCT) facilitates the acquisition of cross-sectional imaging in angiography suites using a rotational C-arm and digital flat panel detectors. The applications are numerous, including evaluation of implanted devices and localization of cerebrovascular lesions. We present and validate the clinical utility of an alternative fast CBCT acquisition protocol in the context of neurovascular device imaging.
Methods:
Contrast-enhanced (CE)-CBCT images were acquired using a new 10 s protocol in a phantom head model, swine model, and in patients. The acquisition parameters of both the 10 s and 20 s protocols were exactly the same, except for fewer projections (250 projections in 10 s vs 500 projections in 20 s), resulting in reduced scan time. Image quality was measured quantitatively in a controlled phantom study and qualitatively by blinded reviewers. The latter was performed to assess the image quality of the 10 s protocol pertinent to the device visibility and its apposition to the parent artery.
Results:
10 s CBCT images were comparable to 20 s CBCT in both phantom and animal studies. Of the 25 patient images, the reviewers agreed that they were able to discern the flow diverter struts and assess the apposition in all images. The overall rating for all 10 s images was 4.28 on a 5-point scale. No images were rated as less than 3, which was the average diagnostic quality. The ratings were concordant across three blinded reviewers (κ=0.411). Additionally, contrast and spatial resolution between 10 s and 20 s images were similar in non-human models.
Conclusions:
CBCT images of neurovascular devices can be obtained successfully using a 10 s acquisition protocol. In addition, the 10 s protocol offers faster acquisition, thus allowing its use in awake patients and with an added advantage of lower radiation and contrast dose.
More Related Videos
05:37Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
10:23Author Spotlight: Three-Dimensional Cephalometric Landmark Annotation Demonstration on Human Cone Beam Computed Tomography Scans
Published on: September 8, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Initiation of Translation
First, the initiator tRNA must be selected from the pool of elongator tRNAs by eukaryotic initiation factor 2 (eIF2). The initiator tRNA (Met-tRNAi) has conserved sequence elements including modified bases at...
Initiation of Translation
Beams
Based on geometry, beams can be straight, tapered, or curved. Straight beams are the most common type and have a constant cross-section throughout their length. Tapered beams, on the other hand, have a varying cross-section along...
Transcription Initiation
The promoters and enhancers and their accessory proteins allow tight regulation of...