Increasing Signal-to-Noise Ratio in Gas Chromatography - Electroantennography Using a Deans Switch Effluent Chopper
Andrew J Myrick1, Thomas C Baker2
1Department of Entomology, Chemical Ecology Laboratory, Penn State University, University Park, PA, 16802, USA. ajm25@psu.edu.
Abstract:
Gas-chromatography-electroantennographic detection (GC-EAD) is a technique used in the identification of volatile organic compounds (VOCs), such as pheromones and plant host odors, which are physiologically relevant to insects. Although pheromones often elicit large EAD responses, other behaviorally relevant odors may elicit responses that are difficult to discern from noise. Lock-in amplification has long been used to reduce noise in a wide range of applications. Its utility when incorporated with GC-EAD was demonstrated previosuly by chopping (or pulsing) effluent-laden air that flowed over an insect antenna. This method had the disadvantage that it stimulated noise-inducing mechanoreceptors and, in some cases, disturbed the electrochemical interfaces in a preparation, limiting its performance. Here, the chopping function necessary for lock-in amplification was implemented directly on the GC effluent using a simple Deans switch. The technique was applied to excised antennae from female Heliothis virescens responding to phenethyl alcohol, a common VOC emitted by plants. Phenethyl alcohol was always visible and quantifiable on the flame ionization detector (FID) chromatogram, allowing the timing and amount of stimulus delivered to the antennal preparation to be measured. In our new chopper EAG configuration, the antennal preparation was shielded from air currents in the room, further reducing noise. A dose-response model in combination with a Markov-chain monte-carlo (MCMC) method for Bayesian inference was used to estimate and compare performance in terms of error rates involved in the detection of insect responses to GC peaks visible on an FID detector. Our experiments showed that the predicted single-trial phenethyl alcohol detection limit on female H. virescens antennae (at a 5.0% expected error rate) was 140,330 pg using traditional EAG recording methods, compared to 2.6-6.3 pg (5th to the 95th percentile) using Deans switch-enabled lock-in amplification, corresponding to a 10.4-12.7 dB increase in signal-to-noise ratio.
Related Concept Videos
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....