CLSI-based transference and verification of CALIPER pediatric reference intervals for 29 Ortho VITROS 5600 chemistry
Victoria Higgins1, Dorothy Truong2, Amy Woroch3
1Department of Laboratory Medicine & Pathobiology, University of Toronto, Toronto, ON M5S 1A8, Canada; CALIPER Program, Pediatric Laboratory Medicine, The Hospital for Sick Children, Toronto, ON M5G 1X8, Canada.
Insights
The CALIPER project successfully transferred pediatric reference intervals (RIs) for 29 biochemical assays to the Ortho VITROS 5600 system. This expands the availability of accurate RIs for pediatric laboratory testing on this platform.
Area of Science:
- Clinical Chemistry
- Laboratory Medicine
- Pediatric Diagnostics
Background:
- Accurate pediatric reference intervals (RIs) are crucial for interpreting laboratory results.
- The Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) developed a pediatric RI database.
- CALIPER RIs were initially established for Abbott ARCHITECT assays.
Purpose of the Study:
- To transfer and validate pediatric reference intervals (RIs) for 29 biochemical assays to the Ortho VITROS 5600 Chemistry System.
- To expand the CALIPER pediatric RI database for use with Ortho analytical platforms.
Main Methods:
- Method comparison analysis using Clinical Laboratory Standards Institute (CLSI) guidelines.
- Measurement of approximately 200 patient serum samples on both Abbott and Ortho assays.
- Linear regression analysis to transfer RIs, followed by verification with 84 healthy pediatric samples.
Main Results:
- Successful transfer and verification of 29 pediatric reference intervals (RIs) for biochemical assays.
- 90% of results from reference samples fell within the transferred confidence limits.
- Calcium, CO2, total bilirubin, magnesium, and LDH did not meet criteria for transference or verification.
Conclusions:
- The CALIPER pediatric RI database is now applicable to laboratories using Ortho VITROS 5600 assays.
- This study enhances the utility of pediatric RIs for clinical laboratories.
- Verification of CALIPER RIs on specific platforms and local populations is recommended.
Background:
Evidence-based reference intervals (RIs) are essential to accurately interpret pediatric laboratory test results. To fill gaps in pediatric RIs, the Canadian Laboratory Initiative on Pediatric Reference Intervals (CALIPER) project developed an age- and sex-specific pediatric RI database based on healthy pediatric subjects. Originally established for Abbott ARCHITECT assays, CALIPER RIs were transferred to assays on Beckman, Roche, Siemens, and Ortho analytical platforms. This study provides transferred reference intervals for 29 biochemical assays for the Ortho VITROS 5600 Chemistry System (Ortho).
Methods:
Based on Clinical Laboratory Standards Institute (CLSI) guidelines, a method comparison analysis was performed by measuring approximately 200 patient serum samples using Abbott and Ortho assays. The equation of the line of best fit was calculated and the appropriateness of the linear model was assessed. This equation was used to transfer RIs from Abbott to Ortho assays. Transferred RIs were verified using 84 healthy pediatric serum samples from the CALIPER cohort.
Results:
RIs for most chemistry analytes successfully transferred from Abbott to Ortho assays. Calcium and CO2 did not meet statistical criteria for transference (r2<0.70). Of the 32 transferred reference intervals, 29 successfully verified with approximately 90% of results from reference samples falling within transferred confidence limits. Transferred RIs for total bilirubin, magnesium, and LDH did not meet verification criteria and are not reported.
Conclusions:
This study broadens the utility of the CALIPER pediatric RI database to laboratories using Ortho VITROS 5600 biochemical assays. Clinical laboratories should verify CALIPER reference intervals for their specific analytical platform and local population as recommended by CLSI.
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Prediction Intervals
However, the point estimate is most likely not the exact value of the population parameter, but close to it. After calculating point estimates, we construct interval estimates, called confidence intervals or prediction intervals. This prediction interval comprises a range of values unlike the point estimate and is a better predictor of the observed sample value, y.
Strategies of Self-Presentation II: Self-Verification
Directing Effect of Substituents: ortho–para-Directing Groups
Confidence Intervals
A...
Inertial Frames of Reference