Paravalvular leakage in patients with prosthetic heart valves: cardiac computed tomography findings and clinical
Hyun Jung Koo1, Joo Yeon Lee1, Gun Ha Kim1
1Department of Radiology and Research Institute of Radiology, Cardiac Imaging Center, Asan Medical Center, University of Ulsan College of Medicine, Asanbyeongwon-gil 86, Seoul, South Korea.
Aims:
Morphological characteristics of paravalvular leakage (PVL) have not yet been well characterized by computed tomography (CT). The purpose of this study was to demonstrate the morphological characteristics of PVLs using cardiac CT in patients diagnosed with PVL.
Methods And Results:
Between May 2011 and December 2013, 46 patients who had been diagnosed with PVL and underwent cardiac CT were included in this study. On CT, the characteristics of PVLs including number, size, location, and shape are described. Inter- and intra-observer agreement of CT were assessed. The extent of PVL on CT and the degree of regurgitant grade on echocardiography are compared. The size of PVLs were compared between patients who underwent surgical correction and who treated with percutaneous device closure or observed without treatment. All PVLs detected on surgical filed were the same with the number and locations of PVLs demonstrated on CT. Interobserver agreement for CT measurements of the PVL was good, ranging from 0.78 to 0.99. The sizes of PVL were smaller in patients who treated with device closure or observed without treatment compared to those in patients who underwent surgical correction (median areas of aortic PVL, 64.5 vs. 15.5 mm2 and mitral PVL, 40.0 vs. 27.0 mm2). PVL sizes were larger in patients with higher regurgitant grades on echocardiography (P < 0.05).
Conclusion:
Cardiac CT can demonstrate the location and size of PVL. The size of PVLs are larger in patients who managed by surgical correction. PVL size measured on CT is correlated with the regurgitant grade on echocardiography.
More Related Videos
Related Concept Videos
Heart Valves
The AV valves prevent the backflow of blood from the ventricles to the atria during ventricular contraction. These valves function with the assistance of the chordae tendineae and papillary muscles. When the ventricles are relaxed, the chordae tendineae are slack, allowing blood to flow from the atria into the...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Toxidromes: Clinical Features
COPD: Pathogenesis and Clinical Features
The primary cause for the onset of COPD is cigarette smoking and exposure to air pollution. These hazardous factors initiate a chain reaction within the lungs, resulting in chronic inflammation, damage to the airways, and a...
Esophageal Strictures-II: Clinical Features and Management
Healthcare providers should gather a comprehensive medical history and conduct a physical examination for diagnosis. If esophageal stricture is...
Endocarditis II: Clinical Features of Infective Endocarditis