A rare case of nonresterilized reinforced ETT obstruction caused by a structural defect: A case report
Sang Ho Kim1, A Yeon Park, Ho Bum Cho
1Department of Anesthesiology and Pain Medicine, Soonchunhyang University Hospital Seoul, Hannam-dong, Yongsan-gu, Seoul, Korea.
Rationale:
Various factors can cause ventilatory failure after endotracheal tube (ETT) intubation, which is associated with increased patient morbidity and mortality.
Patient Concerns:
A 76-year-old woman who was diagnosed with a hemopericardium and suspicion of a major-vessel injury due to dislocation of the clavicular fracture fixation screw.
Diagnosis:
Non-resterilized reinforced ETT obstruction caused by a structural defect.
Intervention:
Endotracheal tube was exchanged.
Outcomes:
The ventilator profile showed rapid improvement.
Lessons:
Anesthesiologists should consider that a non-resterilized reinforced ETT may be defective. An ETT defect can cause high PIP and ETT obstruction without kinking or foreign materials.
Related Concept Videos
Reinforcement
Positive reinforcement occurs when a behavior is followed by the presentation of a rewarding stimulus, increasing the frequency of that behavior. For example:
Corrosion of Reinforcement
However, over time and under certain conditions like carbonation, chloride ingress, and cracking this protective state can be compromised. Steel has areas with...
Reinforcement Schedules
Once a behavior is learned,...
Reinforcements in Concrete
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...
Methods of Sterilization II: Chemical Methods
Using chemical sterilization rather than heat to clean out equipment is recommended. It eradicates and removes all bacteria,...