Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Dose calculation and verification of the Vero gimbal tracking treatment delivery
H Prasetio1, J Wölfelschneider1, M Ziegler1
1Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg, Erlangen, Germany.
Abstract:
The Vero linear accelerator delivers dynamic tumor tracking (DTT) treatment using a gimbal motion. However, the availability of treatment planning systems (TPS) to simulate DTT is limited. This study aims to implement and verify the gimbal tracking beam geometry in the dose calculation. Gimbal tracking was implemented by rotating the reference CT outside the TPS according to the ring, gantry, and gimbal tracking position obtained from the tracking log file. The dose was calculated using these rotated CTs. The geometric accuracy was verified by comparing calculated and measured film response using a ball bearing phantom. The dose was verified by comparing calculated 2D dose distributions and film measurements in a ball bearing and a homogeneous phantom using a gamma criterion of 2%/2 mm. The effect of implementing the gimbal tracking beam geometry in a 3D patient data dose calculation was evaluated using dose volume histograms (DVH). Geometrically, the gimbal tracking implementation accuracy was <0.94 mm. The isodose lines agreed with the film measurement. The largest dose difference of 9.4% was observed at maximum tilt positions with an isocenter and target separation of 17.51 mm. Dosimetrically, gamma passing rates were >98.4%. The introduction of the gimbal tracking beam geometry in the dose calculation shifted the DVH curves by 0.05%-1.26% for the phantom geometry and by 5.59% for the patient CT dataset. This study successfully demonstrates a method to incorporate the gimbal tracking beam geometry into dose calculations. By combining CT rotation and MU distribution according to the log file, the TPS was able to simulate the Vero tracking treatment dose delivery. The DVH analysis from the gimbal tracking dose calculation revealed changes in the dose distribution during gimbal DTT that are not visible with static dose calculations.
Related Concept Videos
Self-Evaluation: Self-Enhancement and Self-Verification
Strategies of Self-Presentation II: Self-Verification
Calculating the Equilibrium Constant
For example, gaseous nitrogen dioxide forms dinitrogen tetroxide according to this equation:
Calculating Standard Free Energy Changes
Calculating pH Changes in a Buffer Solution
Numerical Calculations
The solution to a problem is obtained using different methods. While manually solving algebraic symbols is one of the most common methods, the graphical method is often preferred. Computers...