Water equivalent PRESAGE® for synchrotron radiation therapy dosimetry
Frank M Gagliardi1,2, Liam Day3, Christopher M Poole3
1Alfred Health Radiation Oncology, The Alfred, Melbourne, Vic, 3004, Australia.
A new water equivalent PRESAGE® dosimeter accurately measures 3D dosimetry for synchrotron radiation therapy, validating treatment planning systems. This advancement aids in precise radiation delivery for cancer treatment.
Area of Science:
- Medical Physics
- Radiation Oncology
- Dosimetry
Background:
- Synchrotron Radiation Therapy (SRT) is an emerging cancer treatment modality requiring precise dosimetry.
- Current patient treatment planning systems (TPS) rely on simulated synchrotron x-ray beam data.
- A need exists for a dosimeter capable of accurate 3D dosimetry across SRT energy ranges.
Purpose of the Study:
- To investigate a water-equivalent PRESAGE® dosimeter for 3D dosimetry.
- To evaluate its performance over an energy range suitable for synchrotron x-ray beams.
- To validate its use in conjunction with patient treatment planning systems.
Main Methods:
- Fabricated water-equivalent PRESAGE® dosimeters with a radiological effective atomic number similar to water.
- Irradiated dosimeters across a 10 keV to 10 MeV energy range.
- Performed 3D dosimetry using optical CT (OCT) scanning and compared results with ion chamber measurements and Monte Carlo (MC) simulations.
Main Results:
- PRESAGE® Percentage Depth Dose (PDD) profiles agreed within 2% with ion chamber measurements and MC calculations.
- Good agreement was observed with MC models for depths below 5 mm, where ion chamber data are unavailable.
- While OCT spatial resolution limited penumbra measurement, PRESAGE® successfully verified MC dose profile characteristics.
Conclusions:
- The radiological response of the water-equivalent PRESAGE® dosimeter is validated for SRT x-ray beam energies.
- The dosimeter can independently verify dose distributions generated by MC models.
- This validates PRESAGE® as a tool for accurate 3D dosimetry in SRT applications.