A Modified Technique for Arteriovenous Fistula Construction in Rabbits
Published on: February 10, 2023
Reduction in anastomotic flow disturbance within a modified end-to-side arteriovenous fistula configuration: Results
John Carroll1, Ramon L Varcoe2,3,4, Tracie Barber1
1School of Mechanical and Manufacturing Engineering, University of New South Wales, Australia.
Aim:
We sought to determine if a modification to the peri-anastomotic vein configuration during end-to-side arteriovenous fistula (AVF) creation would achieve a favourable haemodynamic environment compared with the standard acute-angle AVF.
Methods:
Computational fluid dynamics (CFD) modelling of two end-to-side AVF geometries (smooth-vein-loop vs. acute-angle) allowed for haemodynamic modelling. Haemodynamic fields at various stages of the pulse cycle were observed and compared across both models.
Results:
We found a significant reduction in flow disturbance in the modified shape at the peri-anastomotic vein region, a common site for stenosis. The acute anastomotic angle characteristic of a standard-end-to-side AVF was found to cause significant flow disturbance throughout the pulse cycle.
Conclusion:
Computational modelling indicates that improvements in the haemodynamic environment of an AVF are achievable when a gentle change in direction is present in the vein segment. This may lead to improved maturation and AVF patency through reduction in disturbed flow patterns.
Related Concept Videos
Ecological Disturbance
Gene Flow
Flow Cytometry
In...
Flow Sheet
Here's a closer look at the examples of flowsheets commonly used by nurses:
Graphic Sheet Documentation:
Flow Table Test
Concrete is placed within a truncated cone mold that is 8 inches high with an 8-inch base diameter and a 5-inch top diameter. The...
Irrotational Flow