Testing of all Six Semicircular Canals with Video Head Impulse Test Systems
Published on: April 18, 2019
Intra- and Interexaminer Variability of Two Separate Video Head Impulse Test Systems Assessing All Six Semicircular
Emil Riis Abrahamsen1, Ann-Eva Christensen2, Dan Dupont Hougaard1,3
1Department of Otolaryngology, Head and Neck Surgery and Audiology.
Objective:
To evaluate intra- and interexaminer variability of the video Head Impulse Test (v-HIT) when assessing all six semicircular canals (SCCs) of two separate v-HIT systems.
Study Design:
Prospective study.
Setting:
Department of Otolaryngology, Head and Neck Surgery, Aalborg University Hospital, Denmark.
Patients:
One hundred twenty healthy subjects.
Intervention:
Four separate tests of all six SCCs with either system A or system B. Two examiners tested all subjects twice. Pretest randomization included type of v-HIT system, order of paired SCC testing, as well as initial examiner.
Main Outcome Measure:
Gain values and the presence of pathological saccades were registered. Ninety-five percent limits of agreement (LOAs) were calculated for both intra- and interexaminer variability. Adding or subtracting the value from the mean difference achieves the upper and lower bound LOA. Ninety-five percent of the differences lie within these limits.
Results:
Interexaminer reliability: System A: LOAs between 0.13 and 0.24 for the horizontal SCCs and between 0.42 and 0.74 for the vertical SCCs. System B: LOAs between 0.09 and 0.13 for the horizontal SCCs and between 0.13 and 0.20 for the vertical SCCs. Intraexaminer reliability: System A: LOAs were 0.19 and 0.14 for the horizontal SCCs and varied from 0.43 to 0.53 for the vertical SCCs. System B: LOAs were 0.14 for the horizontal SCCs and varied from 0.13 to 0.22 for the vertical SCCs.
Conclusion:
Horizontal SCC testing: both v-HIT systems displayed good intra- and interexaminer variability. Vertical SCC testing: System B displayed good intra- and interexaminer variability whereas the opposite was true with system A.
Related Concept Videos
Systems of Linear Equations in Two Variables
Multi-input and Multi-variable systems
In the absence of...
Impulse
Additionally, it can be shown that the...
Impulse Response
Kirchhoff's law forms an input signal equation, with the capacitor's current and voltage providing the output. Substituting the current and dividing by RC yields a differential equation. The output for an impulse input is the impulse...
Variability: Analysis
The range is a simple measure of variability, indicating the difference between the highest and...
Random Variables
Uppercase letters such as X or Y denote a random variable. Lowercase letters like x or y denote the value of a random variable. If X is a random variable, then X is written in words, and x is given as a number.
For example, let X = the...