Oversensing of transthoracic excitation stimuli in contemporary pacemakers
James D Ryan1, Nathan D Tempel1, Dean D Engle1
1Department of Cardiovascular Medicine, Mayo Clinic, Rochester, MN, USA.
Background:
Transthoracic impedance measurements (TIM) is primarily used for minute ventilation rate adaptive sensors in pacemakers. With elevated impedance, the TIM electrical signal itself is oversensed, causing device malfunction.
Objective:
We report an increased incidence of TIM oversensing.
Method:
Retrospective chart review. We review existing records of 18 patients who have demonstrated device malfunction with TIM oversensing.
Results:
We have found a 1.8% incidence of TIM-related oversensing in our patient population of 959 patients with contemporary Boston Scientific (Marlborough, MA, USA) pacemakers and cardiac resynchronization therapy pacemakers. One patient experienced a syncopal episode.
Conclusion:
Oversensing with pacing inhibition is apparent with the potential of adverse effects to patients.
More Related Videos
06:03Use of Electromagnetic Navigational Transthoracic Needle Aspiration E-TTNA for Sampling of Lung Nodules
Published on: May 23, 2015
12:12Irrelevant Stimuli and Action Control: Analyzing the Influence of Ignored Stimuli via the Distractor-Response Binding Paradigm
Published on: May 14, 2014
Related Concept Videos
Contemporary Psychology
Biopsychology
Biopsychology, also known as biological psychology or behavioral neuroscience, focuses on the biological underpinnings of behavior and mental processes. It...
Excitation-Contraction Coupling in Skeletal Muscles
When an action...
Modified-Release Drug Delivery Systems: Stimuli-Activated
The Quantum-Mechanical Model of an Atom
What is a Sensory System?
Communication